Fluke 96270A 27 GHz Low Phase Noise Reference Source

Manufacturer: Fluke Calibration
Model: 96270A

Features

  • Covers a broad range of RF calibration workload
  • Reduces the number of instruments and interconnections required for your RF calibration system
  • Integrated 300 MHz frequency counter and dual power meter readout eliminate need for additional instruments
  • Calibration-specific interface simplifies technician tasks
  • Simplifies uncertainty calculations
  • Lowers RF system maintenance costs
  • With automation, reduces spectrum analyzer calibration times by as much as 50 % over manual methods

96270A

Specifications

SpecificationFrequencyLevel
Frequency/ Level RangeLeveling Head O/P [50 Ω]: 1 mHz to 4 GHz
Microwave O/P Direct: 1 mHz to 27 GHz [with LL option]
Microwave O/P with HF option: 1 kHz to 27 GHz [with LL option]
-130 to +24 dBm to 125 MHz, +14 dBm at 4 GHz
-4 [-100] to +24 dBm, >1.4 GHz: +20 dBm, [>20 GHz: +18 dBm]
-10 [-35] to +18 dBm, >1.4 GHz: +14 dBm, [>20 GHz: +12 dBm]
Resolution10 µHz0.001 dB
Frequency accuracy± 0.05 ppm ± 5 µHz
Level accuracy (into 50 Ω)Leveling Head O/P:
Down to -48 dBm; ± 0.03 dB to 100 kHz, ± 0.05 dB to 128 MHz, ± 0.3 dB at 4 GHz
10 MHz to 128 MHz; ± 0.05 dB to -48 dBm, ± 0.1 dB to -84 dBm, ± 0.7 dB at -130 dBm
Microwave O/P Direct:
Typically ± 0.5 dB to 4 GHz, ± 0.5 dB to 26.5 GHz
Microwave O/P with HF option (after self-characterization):
Power flatness uncertainty; ± 0.05 dB at 100 MHz, ± 0.07 dB at 1 GHz, ± 0.1 dB at 12 GHz, ± 0.16 dB at 26.5 GHz
Attenuation accuracy (into 50 Ω)Leveling Head O/P: ± 0.02 dB to 49 dB, ± 0.15 dB at 110 dB Relative to +10 dBm, 10 Hz to 128 MHz
VSWRLeveling Head O/P (50 Ω): ≤ 100 MHz: ≤ 1.05, ≤ 2 GHz: ≤ 1.1, 2 GHz to 4 GHz: ≤ 1.0 + 0.05 x f (GHz)
Microwave O/P: direct ≤ 2.0 (typical), HF option splitter device ≤ 1.22
Harmonics and sub-harmonicsHarmonics ≤1 GHz: -60 dBc, >1 GHz: -55 dBc
Sub-harmonics ≤ 4 GHz: none, >4 GHz: -60 dBc
Spurious ≥ 3 kHz offset≤ -84 dBc at 500 MHz, ≤ -78 dBc at 1 GHz, ≤ -66 dBc at 4 GHz, ≤ -48 dBc at 27 GHz
Phase noise at 1 GHz–144 dBc/Hz, typical, at 10 kHz to 100 kHz offset
ModulationAM, FM, PM, internal and external to 4 GHz (Leveling Head and Microwave O/P); frequency pull and external leveling
Frequency sweep1 mHz to 4 GHz (Leveling Head); 1 mHz to 27 GHz (Microwave O/P). Linear or logarithmic. Stop-start or center-span
Frequency counterInternal 300 MHz frequency counter
Power meter readout compatibilityRhode & Schwarz NRP-Z series thermal sensor models 51, 52, 55.03, 55.04, 56, 57, and 58
TemperatureOperating: 0 °C to 50 °C, 23 °C ± 5 °C for specified performance
Storage: -20 °C to +70 °C
Standard interfacesIEEE-488.2 (GPIB)
GPIB command emulation9640A, 9640A-LPN, 9640A-LPNX, HP3335, HP8662A, HP8663A, HP8340A, HP8360 B-Series; Agilent E8257 Series
Dimensions (HxWxD)146 mm x 442 mm x 551 mm (5.8 in x 17.4 in x 21.7 in) including handles. Industry-standard 19 in (483 mm) rack mounting when fitted with Y9600 rack mounting kit
Weight18 kg (40 lb)
*Key specifications summary. Refer to the 96270A Extended Specifications for detailed specifications, including the 75 ohm output Leveling Head.

 