Fluke
Fluke 7009/7108/7015 Resistor Baths
Manufacturer: Fluke Calibration
Model: 7009/7108/7015
Features
- Three size options for any quantity of resistors
- Stability to ±0.0007 °C
- Independent high- and low-temperature cutout circuit
- Minimal long-term drift
Specifications
Specifications |
7015 |
7009 |
7108 |
Range | 0 °C to 50 °C† | 0 °C to 50 °C† | 20 °C to 30 °C |
Stability at 25 °C | ±0.0007 °C (water) ±0.001 °C (mineral oil 5011) |
±0.002 °C (water) ±0.004 °C (mineral oil 5011) |
|
Uniformity | ±0.003 °C at 25 °C (water) ±0.005 °C at 25 °C (mineral oil 5011) |
±0.005 °C (water) ±0.008 °C (mineral oil 5011) |
|
Temperature Setting | Digital display with push-button data entry | ||
Set-Point Resolution | 0.002 °C; high-resolution mode, 0.00003 °C | ||
Display Resolution | 0.01 °C | ||
Digital Setting Accuracy | ±1 °C | ±0.5 °C | |
Digital Setting Repeatability | ±0.01 °C | ||
Heaters | 500 and 1000 Watts | Peltier heating/cooling | |
Cooling Capacity | 100 to 200 Watts | 100 W in ambient 23 °C | |
Access Opening | 699 x 279 mm (27.5 x 11 in) | 699 x 559 mm (27.5 x 22 in) | 356 x 356 mm (14 x 14 in) |
Bath Chamber Dimensions (HxWxD) (unobstructed space) | 699 x 279 x 330 mm (27.5 x 11 x 13 in) |
669 x 559 x 330 mm (27.5 x 22 x 13 in) |
356 x 203 x 355 mm (14 x 8 x 14 in) |
Depth | 330 mm (13 in) | 203 mm (8 in) | |
Wetted Parts | 304 stainless steel | Tank: 304 stainless steel Resistor rack: hard-anodized, perforated aluminum |
|
Safety Cutout | Factory-set high temperature | n/a | |
Power | 115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify | 230 VAC (±10 %), 50 or 60 Hz, 12 A (specify frequency) | 115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify |
Volume | 95 liters (25 gallons) | 167 liters (44 gallons) | 51 liters (13.2 gallons) |
Weight | 141 kg (310 lb.) | 150 kg (330 lb.) | 35 kg (75 lb.) |
Size (HxWxD) | 1219 x 1118 x 559 mm (48 x 44 x 22 in) |
1092 x 1130 x 864 mm (43 x 44.5 x 34 in) |
489 x 413 x 559 mm (19.25 x 22 x 25 in) |
Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration | ||
†Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50°C to protect standard resistors placed in the bath. |