Fluke 6109A / 7109A Portable Calibration Baths

Manufacturer: Fluke Calibration
Model: 6109A / 7109A

Features

  • Calibrate up to four tri-clamp sanitary sensors per batch with ± 0.1 °C temperature display accuracy.
  • Covers a wide temperature range: 6109A: 35 °C to 250 °C
  • 7109A: -25 °C to 140 °C.
  • Stainless steel casing that withstands harsh cleaning solutions; it’s easy to transport and maintain.

6109A / 7109A

Specifications

SpecificationDetails
SizeHeight: 382 mm (15 in)
Width: 242 mm (9.5 in)
Depth: 400 mm (15.7 in)
Weight6109A: 16 kg (35 lb)
7109A: 20 kg (44 lb)
Fluid volume2.5 liters (0.66 gallons)
Fluid working area75 mm x 75 mm (3 in x 3 in)
Maximum fluid depth154 mm (6.1 in)
Factory calibrationTraceable NVLAP-accredited calibration included

Temperature Specifications

The temperature specifications describe the Absolute Instrumental Uncertainty at 95% level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C.

Specification6109A7109A
Range (at 25 °C ambient temperature)35 °C to 250 °C–25 °C to 140 °C (–15 °C cover off)
Accuracy (maximum permissible error)0.1 °C0.1 °C
Display resolution0.01 °, 0.001°0.01 °, 0.001°
Stability0.015 °C0.01 °C
Typical uniformity0.03 °C up to 200 °C
0.04 °C above 200 °C
0.02 °C
Repeatability0.04 °C0.04 °C
Typical heating time35 °C to 100 °C: 25 minutes
100 °C to 250 °C: 45 minutes
–25 °C to 25 °C: 35 minutes
25 °C to 140 °C: 55 minutes
Typical cooling time250 °C to 100 °C: 35 minutes
100 °C to 35 °C: 55 minutes
140 °C to 25 °C: 45 minutes
25 °C to –25 °C: 75 minutes
Typical settling time15 minutes10 minutes

Process Input Module specifications (-P models)

The Input Module specifications describe the Absolute Instrumental Uncertainty at 95% level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C.

FunctionRangeAccuracy (Maximum Permissible Error)
Reference input resistance0 Ω to 42 Ω
42 Ω to 400 Ω
0.0025 Ω
0.006%
Reference input temperature–25 °C to 250 °C0.007% + 0.015 °C
Resistance sensing current1 mA8%
DUT 4-wire resistance0 Ω to 31 Ω
31 Ω to 400 Ω
0.0025 Ω
0.008%
DUT 3-wire resistance accuracy0 Ω to 400 Ω0.12 Ω
Thermocouple mV–10 mV to 100 mV0.025% + 0.01 mV
Reference junction temperature0 °C to 40 °C0.35 °C
Thermocouple temperature–25 °C to 250 °C
0 °C to 250 °C
J: 0.44 °C
E: 0.44 °C
L: 0.42 °C
K: 0.49 °C
N: 0.57 °C
U: 0.48 °C
T: 0.53 °C
M: 0.48 °C
R: 1.92 °C
D: 1.12 °C
S: 1.88 °C
G/W: 3.34 °C
C: 0.84 °C
mA range0 mA to 22 mA0.02% + 0.002 mA
Loop power24 V dc± 6 V

 

REQUEST QUOTATION

PAYMENT

payment-methods

Return Top