Fluke
Fluke 9170 Series Metrology Well Calibrator
Manufacturer: Fluke Calibration
Model: 9170 Series
Features
On-site calibration for lab-level accuracy
- The world’s most accurate, stable and uniform industrial heat source
- Immersion depth: up to 203 mm
- Models with ITS-90 reference temperature probe have an accuracy of ±0.006 °C
- The four models cover a temperature range of –45 °C to 700 °C
Specification
specification | Built-in reference thermometer | ||||
Temperature range | –200 °C 962 °C | ||||
Resistance Range | 0 Ω to 400 Ω, auto-ranging | ||||
Characteristics | ITS-90 Subranges 4, 6, 7, 8, 9, 10, 11 Callendar-Van Dusen (CVD): R 0 , a, b, d | ||||
Resistance Accuracy | 0 Ω to 20 Ω: 0.0005 Ω 20 Ω to 400 Ω: 25 ppm |
||||
Temperature accuracy (not including probe uncertainty) |
10 Ω PRT: ± 0.013 °C at 0 °C 0.014 °C at 155 °C ± 0.019 °C at 425 °C ± 0.028 °C at 700 °C |
25 Ω and 100 Ω PRT: ± 0.005 °C at –100 °C ± 0.007 °C at 0 °C 0.011 °C at 155 °C ± 0.013 °C at 225 °C 0.019 °C at 425 °C ± 0.027 °C at 661 °C |
|||
Resistance Resolution | 0 Ω to 20 Ω: 0.0001 Ω 20 Ω to 400 Ω: 0.001 Ω |
||||
Measurement interval | 1 second | ||||
Connecting to a probe | Shielded 4-wire, 5-pin DIN connector | ||||
Proofreading | NVLAP accredited (built-in reference thermometer only), traceable calibration |