Fluke 9170 Series Metrology Well Calibrator

Manufacturer: Fluke Calibration
Model: 9170 Series

Features

9170 Series

On-site calibration for lab-level accuracy

  • The world’s most accurate, stable and uniform industrial heat source
  • Immersion depth: up to 203 mm
  • Models with ITS-90 reference temperature probe have an accuracy of ±0.006 °C
  • The four models cover a temperature range of –45 °C to 700 °C

Specification

specificationBuilt-in reference thermometer
Temperature range–200 °C 962 °C
Resistance Range0 Ω to 400 Ω, auto-ranging
CharacteristicsITS-90 Subranges 4, 6, 7, 8, 9, 10, 11 Callendar-Van Dusen (CVD): R 0 , a, b, d
Resistance Accuracy0 Ω to 20 Ω: 0.0005 Ω
20 Ω to 400 Ω: 25 ppm
Temperature accuracy
(not including probe uncertainty)
10 Ω PRT:
± 0.013 °C at 0 °C
0.014 °C at 155 °C
± 0.019 °C at 425 °C
± 0.028 °C at 700 °C
25 Ω and 100 Ω PRT:
± 0.005 °C at –100 °C
± 0.007 °C at 0 °C
0.011 °C at 155 °C
± 0.013 °C at 225 °C
0.019 °C at 425 °C
± 0.027 °C at 661 °C
Resistance Resolution0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω
Measurement interval1 second
Connecting to a probeShielded 4-wire, 5-pin DIN connector
ProofreadingNVLAP accredited (built-in reference thermometer only), traceable calibration

REQUEST QUOTATION

PAYMENT

payment-methods