Fluke 9011 High Accuracy Dual Well Calibrator

Manufacturer: Fluke Calibration
Model: 9011

Features

9011

Highest temperature range for a single dry well

  • Two blocks per unit; combined temperature range from –30 °C to 670 °C
  • High temperature block and low temperature block are controlled independently
  • Stability ±0.02 °C
  • Hot block allows for simultaneous calibration of up to eight probes

Specification

specificationHigh temperature blockLow temperature block
range50°C to 670°C–30°C to 140°C
Accuracy± 0.2 °C at 50 °C
± 0.4 °C at 400 °C
± 0.65 °C at 600 °C
± 0.25 °C (insert wells)
± 0.65 °C (fixed wells)
Stability± 0.02 °C at 100 °C
± 0.06 °C at 600 °C
± 0.02 °C at –30 °C
± 0.04 °C at 140 °C
Uniformity± 0.2 °C (typical ± 0.05 °C)± 0.05 °C (insert wells)
± 0.25 °C (fixed wells)
Well Depth152 mm124 mm
Heating time to maximum temperaturehalf an hour15 min
Cooling Time120 min (660 °C to 100 °C)30 minutes (140°C to –30°C)
Insert WellHolds one insert (insert is replaceable)One exchangeable insert well and four fixed wells (3.2 mm, 4.8 mm x2, 6.35 mm)
InterfaceRS-232 interface
power supply115 V ac (± 10 %), 8.8 A or 230 V ac (± 10 %), 4.4 A, switchable, 50/60 Hz, 1150 W
Dimensions (HxWxD)292 x 394 x 267 mm
weight16.4 kg
NIST traceable calibrationData points: 50 °C, 100 °C, 200 °C, 300 °C,
400 °C, 500 °C, 600 °C, 660 °C
Data points: –30 °C, 0 °C, 25 °C, 50 °C, 75 °C,
100 °C, 125 °C, 140 °C