Fluke
Fluke 9011 High Accuracy Dual Well Calibrator
Manufacturer: Fluke Calibration
Model: 9011
Features
Highest temperature range for a single dry well
- Two blocks per unit; combined temperature range from –30 °C to 670 °C
- High temperature block and low temperature block are controlled independently
- Stability ±0.02 °C
- Hot block allows for simultaneous calibration of up to eight probes
Specification
specification | High temperature block | Low temperature block |
range | 50°C to 670°C | –30°C to 140°C |
Accuracy | ± 0.2 °C at 50 °C ± 0.4 °C at 400 °C ± 0.65 °C at 600 °C |
± 0.25 °C (insert wells) ± 0.65 °C (fixed wells) |
Stability | ± 0.02 °C at 100 °C ± 0.06 °C at 600 °C |
± 0.02 °C at –30 °C ± 0.04 °C at 140 °C |
Uniformity | ± 0.2 °C (typical ± 0.05 °C) | ± 0.05 °C (insert wells) ± 0.25 °C (fixed wells) |
Well Depth | 152 mm | 124 mm |
Heating time to maximum temperature | half an hour | 15 min |
Cooling Time | 120 min (660 °C to 100 °C) | 30 minutes (140°C to –30°C) |
Insert Well | Holds one insert (insert is replaceable) | One exchangeable insert well and four fixed wells (3.2 mm, 4.8 mm x2, 6.35 mm) |
Interface | RS-232 interface | |
power supply | 115 V ac (± 10 %), 8.8 A or 230 V ac (± 10 %), 4.4 A, switchable, 50/60 Hz, 1150 W | |
Dimensions (HxWxD) | 292 x 394 x 267 mm | |
weight | 16.4 kg | |
NIST traceable calibration | Data points: 50 °C, 100 °C, 200 °C, 300 °C, 400 °C, 500 °C, 600 °C, 660 °C |
Data points: –30 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, 125 °C, 140 °C |