Fluke 732C/ 734C DC Voltage Standards

Manufacturer: Fluke Calibration
Model: 732C / 734C

Features

  • The Fluke 732C DC Standard is a RoHS compliant direct current voltage standard ideal for maintaining voltages in primary and secondary standard laboratories.
  • The 732C voltage standards provide outputs of 10 V, 1 V, and 0.1 V, respectively, making them ideal for managing standards in the calibration lab or for easy transport to remote locations.
  • The 734C DC voltage standard consists of four electrically and mechanically independent 732C DC standards housed in a rack-sized enclosure.
  • Base and special models are available, with the special model providing twice the stability of the base model.

732C / 734C 732C / 734C

Specifications

Mains power

Table 1 shows the available AC power sources. The AC current draw at 120V AC is 0.13A (typical). Table 1: Mains Power Sources

732C AC voltage settingsUsable AC voltageAvailable frequencies
100V90V to 110V50Hz/60Hz
120V108V to 132V50Hz/60Hz
220V198V to 242V50Hz/60Hz
240V216V to 264V50Hz/60Hz
batteryBattery operationThe product will operate for at least 72 hours on a full charge (at 23 ±5 °C, with a total current drain at the outputs of 0 mA to 0.1 mA).
Charging timeUp to 36 hours (with built-in automatic charger)
External DC InputThe Product can be powered indefinitely through the rear panel input terminals (12 V DC to 15 V DC). The DC power source must be rated ≥300 mA.
insulationThe resistance from any terminal of this product to earth (chassis) ground or to the AC power source is greater than 10,000 MΩ, shunted by less than 1,000 pF.
Guard and Ground TerminalsChassis ground connections are provided on both the front and rear panels, and an internal guard is accessible through a front panel terminal.
Output ProtectionAll outputs can be shorted indefinitely without damage to the unit. The 10 V output can withstand voltages from other sources as follows:For voltages ≤220 V DC, the Product is protected for a maximum continuous current of 50 mA.
For voltages ≤1100 V DC, the Product is protected for short periods against up to 25 mA continuous current or 0.6 Joules.
environment
When operating within specificationsTemperature range: 15 °C to 35 °C
Relative humidity: 0% to 90% up to 28°C, 80% up to 35°C, 50% up to 50°C, non-condensing
Altitude: 0 m to 1830 m (0 ft to 6000 ft)
When operating outside specificationsTemperature range: 0 °C to 50 °C
Relative humidity: 0% to 90%, non-condensing
Altitude: 0 m to 3,050 m (0 ft to 10,000 ft)
During storage (battery removed)Temperature range: -40°C to 50°C
Relative humidity: non-condensing
Altitude: 0m to 12 200m (0ft to 40 000ft)