Okazaki Protection Tube Type Resistance Thermometer Sensors

Manufacturer: Okazaki Manufacturing Company

Specifications

Basic standard

Nominal resistance

ClassTolerance(℃)JIS C1604/IEC 60751ASTM E1137
Winding elementThin film element
Temperature range(℃)
Pt100Ω at 0℃
(R100/R0=1.3851)
AA±〔0.1+0.0017|t|〕-196 to 4500 to 150
A±〔0.15+0.002|t|〕-196 to 450-30 to 300
±〔0.13+0.0017|t|〕-200 to 650
B±〔0.3+0.005|t|〕-196 to 600-50 to 500
±〔0.25+0.0042|t|〕-200 to 650
C±〔0.6+0.01|t|〕-196 to 600-50 to 600
Type
JIS code is in ( ).
Material codeDimension(φ/mm)Max. length
(mm)
Common use temperature
limit(℃)※1
Features
Outer diameterInner diameter
304SS
(SUS304)
A1072000900Excellent at heat/acid/alkali resistance. Weak against sulfur and reducing gas.
129
15113950
316SS
(SUS316)
C1072000900Heat/acid/alkali resistance is the same as the resistance of 304SS; however, corrosion resistance is excellent at high temperatures.
129
15113950
316L SS
(SUS316L)
CL1292000900The amount of C is less than that of 316SS, and the material has grain-boundary corrosion resistance.
15113950
TITANIUMT15113950250Corrosion resistance is extremely excellent at low temperatures; however, the material becomes fragile with oxidation at high temperatures.

※1The common use temperature limits were described referring to JIS and manufacturer’s catalog;
therefore, they are not guaranteed values.