Thomas Scientific Instruments Co.
Thomas Scientific TRL-C40 Circulating Hand Cooler

Manufacturer: Thomas Scientific Instruments Co., Ltd.
Model: TRL-C40
This is a PID-controlled (±0.1℃) cooling water circulation system/chiller for open-air applications. It can circulate cooling water to buckets or your own water tanks. Equipped with various safety devices, it provides constant temperature maintenance with simple operation.
Usage example
Cooling and temperature maintenance by circulation to open containers such as aquariums, flasks, beakers, trays, and ultrasonic aquariums.
Option
The standard circulation pump is a high-head type, but if you require a higher flow rate, you can switch to a 65W magnetic pump (maximum head 3m, maximum flow rate 30L/min). (No additional charge for this change)
Specification
| External dimensions | W405×D570×H775mm |
|---|---|
| Capacity (L) | 8.0 |
| Temperature range (°C) | -20℃ 60℃ When the internal temperature of the tank exceeds 40°C, the refrigerator’s protection circuit will forcibly shut down the refrigerator. Therefore, cooling by the refrigerator is not possible when the internal temperature of the tank is between 60°C and 40°C. |
| temperature stability | ±0.1℃ (Water at 25℃) |
| temperature control | Digital temperature indicator and controller, PID control system |
| timer | Auto-start timer, auto-stop timer (maximum settable time: 99 hours 59 minutes) |
| heater | Sheathed type 1.5kW |
| Refrigeration unit | Air-cooled 400W HFC-404A |
| cooling capacity | 690W (600kcal/h) |
| cooling capacity environment | Water temperature +20℃ 50Hz |
| Circulation pump | 45W magnetic pump, maximum head 8m, maximum flow rate 9L/min, maximum discharge pressure 0.08Mpa |
| Circulation IN-OUT | Outer diameter φ16 hose connector (Rp 1/2) (suitable for hoses with an inner diameter of Φ14 mm) |
| safety equipment | Earth leakage circuit breaker, self-diagnostic function (HC: overheating prevention, LC: low temperature prevention, flow abnormality, refrigerator protection circuit, temperature sensor abnormality, refrigerator pressure abnormality, incorrect setting prevention) |
| power supply | AC100V 22A 50/60Hz, 3m cord (no plug), Power supply capacity 30A |
| others | Drain port outer diameter Φ8mm |
| weight | Approximately 55 kg |
| Price (excluding tax) | ¥1,100,000 |
| option | Circulation pump change: ¥0. This applies only when changing to our designated 65W magnetic pump. |
| Individual Item Catalog | Download individual product catalog (PDF) |
Performance graph

Performance curve
Illustrations

Related Products
-

Thomas Scientific TRL-101FEL Low-temperature Constant-temperature Water Bath
-

Thomas Scientific TRL-140 Low-temperature Constant-temperature Water Bath
-

Thomas Scientific TRL-117V (Non-CFC type) Submersible Hand Cooler
-

Thomas Scientific TRL-107G Submersible Hand Cooler
-

Thomas Scientific TRL-070 Low-temperature Constant-temperature Water Bath
-

Thomas Scientific TRL-130 Low-temperature Constant-temperature Water Bath