Fluke 2456-LEM Laboratory Environmental Monitor

Manufacturer: Fluke Calibration
Model: 2456-LEM

Features

  • Communication via RS-232 or RS-485
  • Includes software for displaying all parameters, recording data and calibration
  • Accuracy: 0.1 °C, 0.04 inHg (1.3 mbar), 3 % relative humidity
  • Ideal for use in calibration labs, clean rooms, storage facilities or with Ruska piston gauges
  • Instead of mechanical chart recorders

Specifications

General Specifications
unitTemperature: °F or °C
Pressure: mmHg, cmHg, inHg, kPa, mbar, psi, bar, kg/cm2
Humidity: %RH, Air Density: g/cm3, kg/m3, lb/in3
power supply Optional power supply via 10 to 36 V dc, 250 mA max (supplied via 115/230 AC adapter) RS-485.
Temperature EnvironmentOperating : 18 °C to 28 °C (64 °F to 82 °F)
Storage: -20 °C to 70 °C (-4 °F to 158 °F)
Humidity Environment5 % to 95 % RH, non-condensing
weight0.23 kg (0.5 lb)
size(H x W x D) 133.35 mm x 82.55 mm x 31.75 mm (5.25 in x 3.25 in x 1.25 in)
communicationRS-232 or RS-485
Warm-up timehalf an hour
Data Update1 second
range
Temperature Environment18°C to 28°C (64°F to 82°F)
pressure20 inHg to 34 inHg (700 mbar to 1150 mbar)
Humidity Environment5 to 95% RH
specification
AccuracyTemperature : ±0.1°C (0.18°F)
Pressure: ±0.04 inHg (1.3 mbar)
Humidity: ±3% RH (Accuracy defined as the combined effects of linearity, repeatability and hysteresis)
Total uncertaintyTemperature: ± 0.5°C (0.9 °F)/year
Pressure: ± 0.08 inHg (2.7 mbar)/year
Humidity: ± 10 % RH/year
Accuracy (uncertainty) formulas follow recommendations from the ISO Guide for Measurement Uncertainty Formulas and include
accuracy, stability, temperature effects and RSS for 2 sigma (95%) of the calibration standards .
softwareRequires Microsoft® Windows® 95 or higher, 1.6 MB free hard disk space,
32 MB RAM, RS-232 or RS-485 interface,
comprehensive help and system manual included, syntax commands included.