Chino IR-FA Series Fiber Type Radiation Thermometer

Made in Japan
Manufacturer: Chino Corporation
Model: IR-FA Series

Thermometer Models

TypesDetecting ElementsModels
For low temperatureCooling type PbSIR-FACR
For medium to high temperatureCooling type PbSIR-FAI
InGaAs SiIR-FAS
Two-color typeInGaAs/InGaAsIR-FAQI
Si/SiIR-FAQS
Si/InGaAsIR-FAQH

Lens Assembly Models

Core diameter 400μm

Lens Assembly Measuring Distance and Measuring DiameterLens Assembly Measuring Distance and Measuring Diameter
φ1mm at 100mmIR-FL0□□□□
φ12mm at 1000mmIR-FL1□□□□
φ5mm at 500mmIR-FL2□□□□
φ2mm at200mmIR-FL3□□□□
φ4mm at200mmIR-FL4□□□□
φ5mm at150mmIR-FL5□□□□
φ20mm at600mmIR-FL6□□□□
φ8mm at1000mmIR-FL8□□□□

Core diameter 800μm (For low temperature)

 

Lens Assembly Measuring Distance and Measuring DiameterModels
φ10 at150mmIR-FL5□□□□
φ40 at600mmIR-FL6□□□□
φ13 at200mmIR-FL7□□□□

Power Supply Unit Model

NameModel
Power supply unitIR-ZFEP

Features

IR-FA series is an optical fiber transmission type radiation thermometer. Measurement light (radiant energy) captured by a size of a pinky, small lens assembly is led to a detecting element stored in a small box type thermometer. With small lens assembly and flexible optical fiber transmission, this thermometer is useful for measurements of a place where direct view is impossible or a narrow place, and a place where inductive interference occurs electrically.

  • Optical fiber can be used without cooling up to 150°C.
  • Optical fiber does not have electrical conduction so it is applicable for induction heating and explosion proof atmosphere.
  • Laser targeting option is available for checking measuring point.
  • Thermometer is compact, light weight, and DIN rail installation available.
  • For low temperature, for medium to high temperature, and two-color type etc. various types are available.

Standard Temperature Range

TypesDetecting ElementsMeasuring RangeLens assembly
Single-color typeCooling type PbS70 to 250°C
100 to 300°C
IR-FL5□, IR-FL6□, IR-FL7□(Core diameter 800μm)
250 to 800°CIR-FL0□, IR-FL1□, IR-FL2□, IR-FL3□, IR-FL4□(Core diameter 400μm)
150 to 500°C
250 to 800°C
IR-FL5□, IR-FL6□(Core diameter 400μm)
300 to 800°CIR-FL8□
InGaAs150 to 450°C
200 to 700°C
250 to 1000°C
300 to 1300°C
IR-FL5, IR-FL6
250 to 1000°C
300 to 1300°C
350 to 1600°C
IR-FL0, IR-FL1, IR-FL2, IR-FL3, IR-FL4, IR-FL8
Si400 to 900°C
500 to 1200°C
600 to 1800°C
700 to 2400°C
IR-FL5, IR-FL6
600 to 1800°C
700 to 2400°C
800 to 3000°C
IR-FL0, IR-FL1, IR-FL2, IR-FL3, IR-FL4, IR-FL8
Two-color typeInGaAs/InGaAs300 to 1200°C
400 to 1500°C
IR-FL5, IR-FL6
400 to 1500°CIR-FL0, IR-FL1, IR-FL2, IR-FL3, IR-FL4
450 to 1500°CIR-FL8
Si/Si800 to 1600°C
1000 to 2000°C
R-FL0, IR-FL1, IR-FL2, IR-FL3, IR-FL4, IR-FL5, IR-FL6
850 to 1600°C
1000 to 2000°C
IR-FL8
Si/InGaAs600 to 1500°C
700 to 2000°C
800 to 2400°C
1000 to 3000°C
IR-FL0, IR-FL1, IR-FL2, IR-FL3, IR-FL4, IR-FL5, IR-FL6, IR-FL8

Specifications

IR-FA Series

Thermometer

ModelsIR-FACIR-FAIIR-FASIR-FAQIIR-FAQSIR-FAQH
Measuring SystemSingle-color typeTwo-color type
Detecting ElementsCooling type PbSInGaASSiInGaAs/InGaAsSi/SiSi/InGaAs
Measuring Wavelength2.0μm1.55μm0.9μm1.55/1.35μm0.85/1.00μm1.55/0.9μm
Accuracy Rating70°C or more, below 300°C: ±4°C
300°C or more, below 500°C: ±5°C
500°Cor more: ±1.0% of measured value
Below 1000°C: ±5°C
1000°C or more, below 1500°C: ±0.5% of measured value
1500°C or more, below 2000°C: ±1% of measured value
2000°C or more: ±2% of measured value
RepeatabilityWithin 2°C0.2°C
Temperature Drift0.2°C/°C0.1°C or 0.015%/°C of measured value, whichever larger0.2°C/°C or 0.02%/°C of measured value, whichever larger
Resolution70°C or more, below 100°C: Approx. 3°C,
100°C or more, below 200°C: Approx. 2°C, 200°C or more: Approx. 0.5°C
0.5°C1.0°C
Response Time0.01s0.04s
Emissivity (Ratio) CompensationEmissivity set value: 1.999 to 0.050 *1Emissivity set value: 1.999 to 0.050Emissivity set value (ratio): 1.999 to 0.050
Signal ModulationDELAY: Average value tracing (smoothing) (Modulation degree 0.0 to 99.9s, 0.1s step can be set freely), Modulation degree 0=REAL,
PEAK: Max. value tracing (Modulation degree 0, 2, 5, 10°C/s selection), Modulation degree 0=Peak hold)
Display SystemLCD 4-digit (Temperature display, parameter display)
Analog Output4 to 20mA DC Isolated output (Load resistance 500Ω or lower)
Accuracy rating: ±0.2% of output range
Output resolution: 0.01% of output range
Output scaling: It can be set freely within the measuring temperature range.
Dummy output: It can be set freely within the range of 0 to 100% of analog output.
Contact Output2-point, select from high, low, high-high, low-low alarms, or error signal
Photo-coupler 30V DC, Max. 50mA
1-point, high (low) alarm or error signal
Photo-coupler 30V DC, Max. 50mA
Contact Input1-point, peak hold or sample hold, dry contact or open collector
Parameter Settings by KeysOperator mode: Setting of emissivity, signal modulation, alarm and others
Engineering mode: Setting of output scaling, ZERO/SPAN, automatic emissivity calculation: input of reference temperature, output correction and others including option-related-parameters
Operator mode: Setting of emissivity, signal modulation, alarm and others
Engineering mode: Setting of output scaling, ZERO/SPAN, automatic emissivity calculation: input of reference temperature, output correction and others including option-related-parameters
Calculation FunctionZERO/SPAN adjustment,
Automatic emissivity calculation *2,
Output correction
ZERO/SPAN adjustment,
Automatic emissivity calculation,
Output correction
ZERO/SPAN adjustment,
Automatic emissivity calculation,
Output correction
Self-DiagnosisThermometer temperature abnormal, parameter error
OptionLaser TargetingBuilt-in semi conductor laser floodlight, laser beam 1mW or less (645nm) class 2, (Built-in semi conductor laser floodlight is not available for High-sensitivity type)
Analog InputInput signal: 4 to 20mA
Select remote emissivity programming or automatic emissivity calculation input of reference temperature.
Input signal: 4 to 20mADC
Select remote emissivity programming or automatic emissivity calculation input of reference temperature.
Communication InterfaceRS-485
Sending of measured data (up to 1-digit below decimal point), and sending/receiving of parameters
Working Temperature Range5 to 40°C0 to 50°C
Rated Power Supply Voltage24V DC(allowable voltage fluctuation range to 28V DC)
Power ConsumptionMax. 15VAMax. 3VA (Approx. 130mA)
ConnectionCramp type no-screw terminals
Installation MethodDIN-rail installation or wall installation
Case/MaterialCopper boardResin
External Dimensions/WeightW140×H110×D65mm, Approx. 1.0kg
(Thermometer only)
W90×H90×D60mm, Approx. 250g (Thermometer only)
CE Marking ConformityConforms to CE marking (EMC directive:EN5501 Group 1 Class A, EN50082-2)

*1 Correction effective ranges are ε≒1.0 to 0.8 for measured temperature 70 to 80°C and ε≒1.0 to 0.6 for 80 to 90°C.

*2 Emissivity (ratio) is calculated automatically by input reference input temperature through key setting or option analog input.

Fiber optics

Core diameter 400μmCore diameter 800μm (For low temperature type only)
FiberSingle core quartz fiber
Sheath
  • Without metal protective tube (Heat-resistive sheath/glass wool braided)
  • With metal protective tube (Heat-resistive sheath/glass wool braided+SUS flexible tube)
Working Temperature0 to 150°C
  • At measuring temperature 70 to 120°C: 0 to 50°C
  • At measuring temperature 120°C or more: 0 to 80°C
LengthFor low temperature type・・・2m, 4m, 5m
For medium to high temperature・Two-color types・・・2m, 4m, 5m, 10m, 15m, 20m
2m
Allowable Bending RadiusR100mmR150mm