Fluke 5322A Electrical Safety Tester Calibrator

Manufacturer: Fluke Calibration
Model: 5322A

Features

  • Continuously variable high voltage resistance outputs source high-ohms, high voltage resistors to enable calibration of insulation resistance testers/megohmmeters up to 5 kV.
  • The continuously variable feature lets the user set any resistance value to match the device under test (DUT) requirement and adjust the 5322A output to a nominal value, which aids in calibrating older-style hand-cranked testers.
  • Active Loop Compensation makes it easier to calibrate the loop/line impedance function of an installation tester by canceling residual line impedance during calibrations.
  • 600 V source (VLC option) allows calibration of ac voltage measurements on workloads with measurement capabilities. This feature is also useful for powering many types of PAT testers.
  • 4-wire low resistance sources enable precision low current measurements and high current ground bond measurements, providing accuracy for testing newer 4-wire testers.
  • Optional 5322A-LOAD 5 kV high voltage load resistors include paralleling capability for precision hipot testing.
  • RCD simulation enables testing of installation and PAT testers with excellent time and current level accuracy. 5322A trip times provide better than 4:1 test uncertainty ratios with 1 % trip current uncertainty.
  • Ripple coefficient and THD measurements display hipot testers’ signal purity measurements to 5 kV as required by many regulations.
  • Multiple models give you the flexibility to select the capabilities best suited to your workload and budget.
  • MET/CAL software compatibility automates the 5322A in 5320A emulation mode for improved throughput and greater consistency.

5322A

Specifications

Specifications confidence level99%
Specifications interval1 year
Power line115/230 V ac (50/60 Hz) +10 % / -14 %, with the maximum voltage difference between neutral and protective earth not exceeding 15 V. Operation with power line between -10 % and -14 % has limitations in burden current for voltage outputs. See AC/DC Voltage Calibrator (VLC option) below.
Power consumption1250 VA maximum
Fuse protection
  • AC mains input: 2 A, 250 V for 230 V, time delay (T2L250 V – 5 mm x 20 mm) 4 A, 250 V for 115 V, time delay (T4L250 V – 5 mm x 20 mm)
  • RCD input: 3.15 A, 250 V, fast (F3.15H250 V – 5 mm x 20 mm)
  • Meter amps (A) input: 20 A, 500 V, time delay (F20H500 V – 6.3 mm x 32 mm)
  • Loop/Line impedance input: 4 A, 500 V, time delay (T4H500 V – 6.3 mm x 32 mm)
  • Leakage current input: 100 mA, 250 V, fast (F100 mL250 V – 5 mm x 20 mm)
Environment
Warm-up time15 minutes
Temperature performance
  • Operating temperature: 18 °C to 28 °C
  • Calibration temperature (tcal): 23 °C
  • Temperature coefficient for temperatures outside of tcal ± 5 °C is 0.1 / °C of the specification, for temperatures between 5 °C to 40 °C.
  • Storage temperature : -10 °C to 50 °C
  • Storage recovery time: Typically <24 hours at operating temperature
Relative humidity (operating)<80 % to 28 °C (resistance outputs >10 GΩ specified for <70 % to 28 °C)
Relative humidity (storage)90 % non-condensing 0 °C to 50 °C
Altitude

Operating: 2000 m (6561 ft)

Storage: 12 200 m (40 000 ft)