Fluke 5699 High-Temperature Metal-Sheath SPRT

Manufacturer: Fluke Calibration
Model: 5699

Features

  • Range to aluminum point (660 °C )
  • Inconel™ and platinum sheaths guard against contamination
  • Less than 8 mK/year drift
  • Many calibration options available including by fixed point
  • Fifth wire provides shield ground

Specifications

SpecificationDetails
Temperature Range–200 °C to 661 °C (recommended annealing temperature: 665 °C)
Nominal RTPW25.5 Ω (± 0.5 Ω)
Current1 mA
Resistance RatioW(302.9146 K) ≥ 1.11807
W(234.3156 K) ≤ 0.844235
Sensitivity0.1 Ω/°C
Drift Rate< 0.008 °C/year
(< 0.003 °C/year typical)
Repeatability< 1 mK
Self-heating at TPW< 0.001 °C under 1 mA current
Reproducibility± 0.001 °C or better
RTPW Drift After Thermal Cycling< 0.001 °C
Diameter of Pt Sensor Wire0.07 mm (0.003 in)
Lead WiresFour sensor wires plus grounding wire
Protective SheathInconel
Diameter5.56 mm ± 0.13 mm (0.219 in ± 0.005 in)
Length482 mm (19 in)
Insulation Resistance> 100 MΩ at 661 °C
> 1000 MΩ at 20 °C

 