Fluke
Fluke 6054/6055/7007 Deep-Well Baths
Manufacturer: Fluke Calibration
Model: 6054/6055/7007
Features
- Constant liquid levels through concentric-tube design
- Special design for sighting LIG thermometers
- Depth up to 60 cm (24 in)
- Optional interface packages control all settings
Specifications
7007 |
6054 |
6055 |
|
Range | –5 °C to 110 °C | 50 °C to 300 °C | 200 °C to 550 °C |
Temperature Stability | ± 0.001 °C at 0 °C (water/ethanol) ± 0.001 °C at 25 °C (water) 0.003 °C at 100 °C (oil 5012) |
± 0.003 °C at 100 °C (oil 5012) ± 0.005 °C at 300 °C (oil 5017) |
± 0.003 °C at 200 °C (salt) ± 0.01 °C at 550 °C (salt) |
Temperature Uniformity | ± 0.004 °C at 0 °C (ethanol) ± 0.007 °C at 100 °C (oil 5012) |
± 0.007 °C at 100 °C (oil 5012) ± 0.015 °C at 300 °C (oil 5017) |
± 0.005 °C at 200 °C (salt) ± 0.010 °C at 550 °C (salt) |
Temperature Setting | Digital display with push-button data entry | ||
Set-Point Resolution | 0.002 °C, high res. 0.00003 °C | 0.01 °C, high res. 0.00018 °C | |
Display Temperature Resolution | 0.01 °C | ||
Digital Setting Accuracy | ± 1 °C | ||
Digital Setting Repeatability | ± 0.005 °C | ± 0.01 °C | |
Heaters | 250 to 1000 W | 250 to 1000 W | 260 to 2080 W |
Working Area | 178 mm dia. (7 in) | 196 mm dia. (7.7 in) | 107 mm dia. (4.2 in) |
Depth | 24 in | 24 in | 24 in |
Wetted Parts | 304 stainless steel | ||
Power | 230 V ac (± 10 %), 50 or 60 Hz, 15 A (Specify frequency, contact Fluke Calibration if CE mark required.) |
230 V ac (± 10 %), 50/60 Hz, 15 A max |
230 V ac (± 10 %), 50/60 Hz, 15 A |
Volume | 42 liters (11.2 gal) | 50 liters (13.2 gal) | 19.8 liters (5.2 gal, 43 kg [95 lb] of bath salt) |
Size (D x W x H) | 470 x 775 x 194 mm to working surface (18.5 x 30.5 x 47 in), 1397 mm (55 in) to top of stir motor, 914 mm (36 in) to control panel | 572 x 762 x 1219 mm to working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top of stir motor box, 914 mm (36 in) to control panel | 572 x 775 x 1219 mm to working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to top of stir motor box, 914 mm (36 in) to control panel |
Distance from Line of Sight to Top of Fluid | 9.5 mm (3/8 in) | 15.9 mm (5/8 in) | |
Automation | Use the 1586A Super-DAQ to automate temperature sensor calibration. For IEEE-488, add the 2001-IEEE to the automation package. | ||
Weight | 70.8 kg (156 lb) |