Fluke 7009/7108/7015 Resistor Baths

Manufacturer: Fluke Calibration
Model: 7009/7108/7015

Features

  • Three size options for any quantity of resistors
  • Stability to ±0.0007 °C
  • Independent high- and low-temperature cutout circuit
  • Minimal long-term drift

Specifications

Specifications

7015

7009

7108

Range0 °C to 50 °C0 °C to 50 °C20 °C to 30 °C
Stability at 25 °C±0.0007 °C (water)
±0.001 °C (mineral oil 5011)
±0.002 °C (water)
±0.004 °C (mineral oil 5011)
Uniformity±0.003 °C at 25 °C (water)
±0.005 °C at 25 °C (mineral oil 5011)
±0.005 °C (water)
±0.008 °C (mineral oil 5011)
Temperature SettingDigital display with push-button data entry
Set-Point Resolution0.002 °C; high-resolution mode, 0.00003 °C
Display Resolution0.01 °C
Digital Setting Accuracy±1 °C±0.5 °C
Digital Setting Repeatability±0.01 °C
Heaters500 and 1000 WattsPeltier heating/cooling
Cooling Capacity100 to 200 Watts100 W in ambient 23 °C
Access Opening699 x 279 mm (27.5 x 11 in)699 x 559 mm (27.5 x 22 in)356 x 356 mm (14 x 14 in)
Bath Chamber Dimensions (HxWxD) (unobstructed space)699 x 279 x 330 mm
(27.5 x 11 x 13 in)
669 x 559 x 330 mm
(27.5 x 22 x 13 in)
356 x 203 x 355 mm
(14 x 8 x 14 in)
Depth330 mm (13 in)203 mm (8 in)
Wetted Parts304 stainless steelTank: 304 stainless steel
Resistor rack: hard-anodized, perforated aluminum
Safety CutoutFactory-set high temperaturen/a
Power115 VAC (±10 %), 60 Hz, 15 A or 230 VAC, 50 or 60 Hz, 8 A, specify230 VAC (±10 %), 50 or 60 Hz, 12 A (specify frequency)115 VAC (±10 %), 50/60 Hz, 3 A or 230 VAC (±10 %), 50/60 Hz, 1.6 A, specify
Volume95 liters (25 gallons)167 liters (44 gallons)51 liters (13.2 gallons)
Weight141 kg (310 lb.)150 kg (330 lb.)35 kg (75 lb.)
Size (HxWxD)1219 x 1118 x 559 mm
(48 x 44 x 22 in)
1092 x 1130 x 864 mm
(43 x 44.5 x 34 in)
489 x 413 x 559 mm
(19.25 x 22 x 25 in)
AutomationUse the 1586A Super-DAQ to automate temperature sensor calibration
Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with standard resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50°C to protect standard resistors placed in the bath.

REQUEST QUOTATION

PAYMENT

payment-methods

Return Top