Manufacturer: Fluke Calibration
Model: 714B
Features
- The 714B can measure and simulate 17 different thermocouples and mV
- Measures 4 to 20 mA signals and simultaneously emits a temperature signal
- Magnetic hangers designed to fit each device are included as standard.
- Configurable 0% and 100% source settings for quick checking of 25% linearity
- Linear ramp and 25% step auto ramp based on 0% and 100% settings
- Dual input and backlit display for easy reading of measurements
- Settings from when the power was turned off are remembered when the power is turned on, making it easy to resume testing
- 1 year/2 year specifications and traceable calibration certificate
Specifications
General Specifications |
Maximum voltage applied between any terminal and ground, or between any two terminals |
30V |
Operating temperature |
-10°C to 50°C |
Storage temperature |
-30°C to 60°C |
Operating altitude |
2,000m |
Storage altitude |
12,000 m |
Relative Humidity (% RH, non-condensing) |
No condensation |
90% (10°C to 30°C) |
75% (30°C to 40°C) |
45% (40°C to 50°C) |
(No condensation) |
vibration |
MIL-T-28800E, Class 2 |
Drop Test |
1 m |
IP Rating |
IEC 60529: IP52 |
Electromagnetic environment |
IEC 61326-1, portable |
Safety |
IEC 61010-1, 30 V max (to earth), Pollution Degree 2 |
power supply |
4 AA batteries (NEDA 1.5A, IEC LR6) |
Dimensions (H x W x L) |
52.5mm x 84mm x 188.5mm |
weight |
515 g |
DC mA measurement |
resolution |
0 to 24mA |
range |
0.001mA |
Accuracy (% of reading + counts) |
0.010% + 2 μA |
Temperature coefficient |
±(0.002% of reading + 0.002% of range)/°C (<18°C, or >28°C) |
Millivolt Measurement and Source |
resolution |
-10mV to 75mV |
range |
0.01mV |
Accuracy (% of reading + counts) |
0.015% + 10 μA |
Temperature coefficient |
±(0.002% of reading + 0.002% of range)/°C (<18°C, or >28°C) |