Fluke 714B Thermocouple Calibrator

Manufacturer: Fluke Calibration
Model: 714B

Features

  • The 714B can measure and simulate 17 different thermocouples and mV
  • Measures 4 to 20 mA signals and simultaneously emits a temperature signal
  • Magnetic hangers designed to fit each device are included as standard.
  • Configurable 0% and 100% source settings for quick checking of 25% linearity
  • Linear ramp and 25% step auto ramp based on 0% and 100% settings
  • Dual input and backlit display for easy reading of measurements
  • Settings from when the power was turned off are remembered when the power is turned on, making it easy to resume testing
  • 1 year/2 year specifications and traceable calibration certificate

714B

Specifications

General Specifications
Maximum voltage applied between any terminal and ground, or between any two terminals30V
Operating temperature-10°C to 50°C
Storage temperature-30°C to 60°C
Operating altitude2,000m
Storage altitude12,000 m
Relative Humidity (% RH, non-condensing)No condensation
90% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
(No condensation)
vibrationMIL-T-28800E, Class 2
Drop Test1 m
IP RatingIEC 60529: IP52
Electromagnetic environmentIEC 61326-1, portable
SafetyIEC 61010-1, 30 V max (to earth), Pollution Degree 2
power supply4 AA batteries (NEDA 1.5A, IEC LR6)
Dimensions (H x W x L)52.5mm x 84mm x 188.5mm
weight515 g
DC mA measurement
resolution0 to 24mA
range0.001mA
Accuracy (% of reading + counts)0.010% + 2 μA
Temperature coefficient±(0.002% of reading + 0.002% of range)/°C (<18°C, or >28°C)
Millivolt Measurement and Source
resolution-10mV to 75mV
range0.01mV
Accuracy (% of reading + counts)0.015% + 10 μA
Temperature coefficient±(0.002% of reading + 0.002% of range)/°C (<18°C, or >28°C)