Features
The 8270A and 8370A are auto pneumatic high-pressure controllers that calibrate a wide workload of pressure sensors, covering twice the pressure ranges at twice the speed of other high-pressure controllers. Two models let you balance price and performance:
- The 8270A measures and controls pressures from vacuum to 44 MPa (6400 psi). It can configured with ranges as low as 100 kPa (15 psi).
- The 8370A measures and controls pressures from atmosphere to 107 MPa (15 500 psi). It can configured with ranges as low as 700 kPa (100 psi).
Specifications
General specifications | |
Mains | |
Power requirements | 100 V ac to 240 V ac, 47 Hz to 63 Hz |
Fuse | T 2 A, 250 V |
Max power consumption | 100 W |
Environment | |
Operating ambient temperature range | 15 °C to 35 °C |
Storage temperature | -20 °C to 70 °C |
Relative humidity | Operating: <80 % to 30 °C, <70 % to 40 °C |
Storage: <95 %, non-condensing. A power stabilization period of four days may require after extended storage at high temperature and humidity. | |
Vibration | MIL-T-28800E |
Altitude (operation) | <3000 m |
Warm up time | 15 minutes after power up or module installation, when items previously stored within Operating Ambient Temperature Range. |
Compliance | |
Ingress protection | IEC 60529:IP20 |
Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
Weight and Dimensions | |
Weight (Chassis only) | 13 kg (28.5 lbs) / 15 kg (35.25 lbs) |
Dimensions | Height: 147 mm (5.78 in) |
Width: 452 mm (17.79 in) | |
Depth 488 mm (19.2 in) | |
Rack mount dimensions | 3U-19 inch rack |
Electromagnetic Compatibility (EMC) | |
IEC 61326-1 (Controlled EM environment | IEC 61326-2-1; CIPR 11: Group 1, Class A Group 1 equipment has intentionally generated and/or uses conductivity coupld radio-frequency energy which is necessary for the internal functioning of the equipment itself.Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. Emissions which the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of 61326-1 when test leads and/or test probes are connect |
USA (FCC) | 47 CFR 15 subpart B, this product is considered an exempt device per clause 15.103. |
Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intend for use in business environments and not to use in homes. |
Communication Interfaces | |
Primary Remote Interfaces | IEEE, Ethernet, RS-232, USB |
System Connection | Supports interconnection of 2 or 3 systems |
Switch Test Connection | Standard BNC Jack: Nominal 24 V dc isolated drive Maximum 30 V dc w.r.t.chassis ground |
Aux Drivers | 4 external Solenoid Drivers 24 V dc Drive (Maximum drive 6 W continuous per channel |