Fluke 908/909 Frequency Standard

Manufacturer: Fluke Calibration
Model: 908/909

Features

  • Accurate reference “atomic clock” for automated test systems
  • Affordable and extremely economical
  • Optional carrying case for portability
  • No battery backup required during transport
  • Short warm-up time

908/909

Specifications

ParameterDetails
Basic model5×10 MHz, 1×5 MHz: Sine wave >0.6V rms, 50Ω
With option 7010×10 MHz, 1×5 MHz: Sine wave >0.6V rms, 50Ω
Frequency Stability (1 month)908: 3×10⁻⁹, 909: 5×10⁻¹¹
Frequency Stability (1 year)908: < 2×10⁻⁸ (1 year), 909: 1×10⁻⁹ (10 years)
Temperature Environment (20°C – 26°C)908: < 4×10⁻¹⁰ typical, 909: 2×10⁻¹¹ typical value
Temperature Environment (0°C – 50°C)908: 2.5×10⁻⁹, 909: 3×10⁻¹⁰
Short-term stability (10s)908: 5×10⁻¹², 909: 1×10⁻¹¹
Short-term stability (1s)908: 5×10⁻¹², 909: 3×10⁻¹¹
Warm up time (+25°C)908: 10 minutes to 5×10⁻⁹, 909: 5.4 minutes to lock
Operating temperature0°C to +50°C
Storage temperature-40°C to +70°C
Safety standardsEN 61010-1, CE compliant
EMI standardsEN 55011 ISM group, class B, EN 50082-2, CE compliant
Power consumption (908)<20W during warm-up, <7W continuous
Power consumption (909)<70W during warm-up, <30W continuous
Dimensions315 x 86 x 395 mm
Weight4.8 kg (unit only), 7.8 kg (shipping)