Fluke
Fluke 9100S/9102S Handheld Dry Well Calibrator
Manufacturer: Fluke Calibration
Model: 9100S/9102S
Features
The world’s smallest, lightest, and most portable dry-well
- The world’s smallest dry well
- Temperature range: –10°C to 375°C
- Accuracy: ±0.25°C, Stability at 0°C: ±0.05°C
- RS-232 interface
Specification
specification | 9100S | 9102S |
Temperature range | 35°C to 375°C | –10 °C to 122 °C (at 23 °C ambient) |
Accuracy | ± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C | ±0.25°C |
Stability | ± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C | ±0.05°C |
Well-to-well uniformity | ±0.2 °C at similar size sensors and same depths | |
Heating time | 35°C to 375°C: 9.5 minutes | Ambient to 100 °C: 10 minutes |
Stabilization Time | 5 mins | 7 mins |
Cooling Time | 375 °C to 100 °C: 14 min | Ambient to 0 °C: 10 minutes |
Well Depth | 102 mm; 89 mm for 1.6 mm diameter wells | 102 mm |
Removable inserts | none | 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included standard] |
power supply | 115Vac (± 10%), 55–65Hz, 1.5A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz 175 W |
94-234Vac (± 10%), 50/60Hz, 60W
|
Dimensions (HxWxD) | 57 x 125 x 150 mm | 99 x 140 x 175 mm |
weight | 1 kg | 1.8 kg |
Interface | RS-232 interface | |
NIST traceable calibration | Data for 50 °C, 100 °C, 150 °C, 200 °C, 250 °C, 300 °C, 375 °C | Data for –10 °C, 0 °C, 25 °C, 50 °C, 75 °C, 100 °C, 122 |