Fluke
Fluke 9132/ 9133 Portable Infrared Calibrators
Manufacturer: Fluke Calibration
Model: 9132/ 9133
Features
- Certify IR pyrometers from –30 °C to 500 °C (–22 °F to 932 °F)
- Large 57 mm (2.25 in) blackbody target
- RTD reference well for contact temperature measurement
- Small, compact design in a blackbody calibrator
Specifications
Specifications | 9133 | 9132 |
Temperature Range | –30 °C to 150 °C at 23 °C ambient (–22 °F to 302 °F at 73 °F ambient) |
50 °C to 500 °C (122°F to 932°F) |
Type of Calibration | NIST traceable contact calibration | NIST traceable radiometric |
Accuracy | ± 0.4 °C (± 0.72 °F) |
0.5 °C + 0.3 % of reading
0.8 °F + 0.3 % of reading
|
Stability | ± 0.1 °C (± 0.18 °F) | ± 0.1°C at 100°C (± 0.18°F at 212°F) ± 0.3°C at 500°C (± 0.54°F at 932°F) |
Target Size | 57 mm (2.25 in) | |
Target Emissivity |
0.95 (± 0.02 from 8 to 14 μm)
|
Radiometrically calibrated to 0.95 |
Resolution | 0.1 ° | |
Heating Time | 15 minutes (25 °C to 150 °C) | 30 minutes (50 °C to 500 °C) |
Cooling Time | 15 minutes (25 °C to –20 °C) | 30 minutes (500 °C to 100 °C) |
Computer Interface | RS-232 interface | |
Power | 115 V ac (± 10 %), 1.5 A, or 230 V ac (± 10 %), 1.0 A, switchable, 50/60 Hz, 200 W |
115 V ac (± 10 %), 3 A or 230 V ac (± 10 %), 1.5 A, switchable, 50/60 Hz, 340 W |
Size (HxWxD) |
152 x 286 x 267 mm (6 x 11.25 x 10.5 in) |
102 x 152 x 178 mm (4 x 6 x 7 in ) |
Weight | 4.6 kg (10 lb) | 1.8 kg (4 lb) |