Fluke
Fluke Mini Fixed-Point Cells
Manufacturer: Fluke Calibration
Product name: Mini Fixed-Point Cells
Features
- Lower uncertainties than comparison calibrations
- All ITS-90 fixed points from TPW to copper
- Reduced equipment and annual recalibration costs
Specifications
Model Number | Fixed Point | Temperature (°C) | Outside Diameter | Inside Diameter | Total Cell Height | Immersion Depth1 | Cell Only2 | Simple Realization2 |
---|---|---|---|---|---|---|---|---|
5901B-G | Water T.P. | 0.01 | 30 mm | 8 mm | 170 mm | 117 mm | 0.2 | 0.5 |
5914A | Indium F.P. | 156.5985 | 43 mm | 8 mm | 214 mm | 140 mm | 1.0 | 2.0 |
5915A | Tin F.P. | 231.928 | 43 mm | 8 mm | 214 mm | 140 mm | 1.4 | 3.0 |
5916A | Zinc F.P. | 419.527 | 43 mm | 8 mm | 214 mm | 140 mm | 1.6 | 4.0 |
5917A | Aluminum F.P. | 660.323 | 43 mm | 8 mm | 214 mm | 140 mm | 4.0 | 10.0 |
5918A | Silver F.P. | 961.78 | 43 mm | 8 mm | 214 mm | 140 mm | 7.0 | n/a |
5919A | Copper F.P. | 1084.62 | 43 mm | 8 mm | 214 mm | 140 mm | 15.0 | n/a |
5944 | Indium F.P. | 156.5985 | 41.3 mm | 7.8 mm | 222 mm | 156 mm | 0.7 | 1.4 |
5945 | Tin F.P. | 231.928 | 41.3 mm | 7.8 mm | 222 mm | 156 mm | 0.8 | 1.6 |
5946 | Zinc F.P. | 419.527 | 41.3 mm | 7.8 mm | 222 mm | 156 mm | 1.0 | 2.0 |
5947 | Aluminum F.P. | 660.323 | 41.3 mm | 7.8 mm | 222 mm | 156 mm | 2.0 | 4.0 |
Uncertainty (mK) k=2
1 Distance from the bottom of the central well to the surface of the pure metal. 2 “Cell Only” refers to the expanded uncertainty of the cell when realized by traditional methods and maintained using traditional maintenance devices. “Simple Realization” refers to the expanded uncertainty of the cell when realized using practical methods (melting points instead of freezing points or slush ice instead of an ice mantle, for example) and maintained using Hart’s 9210 and 9260 mini cell maintenance apparatus. |