- Home
- Products
- Kyowa DPM-42B-I Dynamic Strain Measurement Card, Inverter Noise Reduction Type EDX Series Conditioner Card
Japan, Kyowa Electronic Instruments Co., Ltd.
Kyowa DPM-42B-I Dynamic Strain Measurement Card, Inverter Noise Reduction Type EDX Series Conditioner Card
Made in Japan
Manufacturer: Kyowa Electronic Instruments Co., Ltd.
Model: DPM-42B-I Dynamic Strain Measurement Card, Inverter Noise Reduction Type
- Conditioner card for EDX-200A and EDX-5000A
Feature
Can be selected according to the application. In addition to strain, voltage, acceleration, and temperature measurements, CAN data is also supported. Noise-resistant isolation amplifier cards and dynamic strain measurement cards are also available.
Specification
Measurement object | Strain gauge, strain gauge transducer |
---|---|
Number of input channels | 4 |
Response frequency range | DC~5kHz (deviation: within ±10%) |
Compatible bridge resistance | 120~1000Ω |
Gauge rate | 2.00 Fixed |
Bridge Power Supply | 2Vrms,0.5VrmsSwitching, 12kHz sine wave |
Equilibrium adjustment range | Resistance : Within ±2.4%(±12000×10-6 Strain) Capacity : Within 2000pF |
Equilibrium adjustment method | Resistance : Pure electronic auto-balance capacity : CST type (automatic tracking) |
Measuring range | Bridge power supply 2VrmsHours: 200,500,1K,2K,5K,10K,20K,10K×<>-6Strain and OFF 8-stage
bridge supply 0.5VrmsHours: 1k,2k,5k,10k,20k,50k×10-6Strain and OFF 7-stage |
Calibration value (CAL) | Output 100% ± and 50% ± of each range |
Nonlinearity | ± within 0.2%FS |
SNR ratio | 44dBp-pOver (500×10-6When strain range is set) |
Lowpass Filter | Transfer characteristics: Second-order Butterworth type cut-off frequency: 2,10,30,100,300k Amplitude ratio at 1-stage cut-off point of FLAT at Hz: -6±3dB Attenuation characteristics: – 1±12dB/oct. |
resolution | 16-bit |
Additional features | Input check function, TEDS sensor compatible |
Monitor output | Accuracy: ±5V±0.5% (at ±FS) Nonlinearity: Within 0.5%FS |
Dielectric voltage | Input-to-output: AC 250V, 1 minute |
Noise immunity | Inverter noise reduction |
Conformity directives | EMC Directive 2014/30/EU RoHS Directive 2011/65/EU, (EU ) 2015/863 (10 substances) |
Sold separately | Monitor output cable U-64Note ) When using a converter with remote sensing, connect it using a 4-core extension cable (N81~N85) for measurement. However, the remote sensing function cannot be used. |
---|