Model Name | EDX-5000A-64-H (HDD)EDX-5000A-64-S (Disk with Internal Storage Unit: SSD) |
---|---|
Compatible conditioners | CDV-40B(-F),CVM-41A,DPM-42B(-F),DPM-42B-I(-F),CTA-40A,CFV-40A,CCA-40A(-F),CDV-44AS,CDA-44AS,CDA-45AS,AD-40AS(-F)*Up to two CFV-40A sheets can be mounted. |
Number of input channels | Up to 64 channels (with 40 CDV-8B devices) |
Analog Input | For details, refer to the specifications of each conditioner card. |
Digital input output
Number of input/output points | Up to 8 points |
---|---|
Input/output settings | Digital input/digital output/remote control input can be switched for each bit (common is common) * Remote control input: Measurement start/stop, BAL execution, etc. are possible. |
Input Format | Isolated, TTL Level Input |
Input voltage | DC max 5V |
Insulation method | Digital isolator |
Output format | Isolated, open-collector output with 10kΩ internal pull-up resistor |
Output voltage | DC 5V |
Output current | Up to 1mA (per point) |
Voice input | 1-channel (voice memos entered during acquisition can be acquired together with measurement data) using RCU-42A remote control unit (optional) |
---|
sampling frequency
Sampling method | 1-2-5 series 1Hz~200kHz: 32Hz~1kHz when acquiring up to 100 channels: 64Hz~1kHz when acquiring up to 50 channels: 80Hz~1kHz whenacquiring up to 10 channels: When real-time simultaneous processing is enabled 2nSeries 2Hz~131.072kHz: 32Hz~2.65kHz when acquiring up to 536 channels: 64Hz~2.32kHz when acquiring up to 768 channels: 80Hz~2.8kHz when acquiring up to 192 channels: When * AD data format: In 16-bit mode * In |
---|
storage | Hard disk drive (HDD) Capacity 300 GB or more Solid state drive (SSD) Capacity 30 GB or more |
---|---|
indication | Channel status LED (input OVER value level can be set for each channel) REC/BUSY/BATTERY/POWER LED 12.1-inch wide touch panel display |
Operation keys | Touch panel and main unit keys (REC/STOP/BAL./OPT.) can also be operated with an external keyboard and mouse |
External Control Connector | CONT. IN,CONT. OUT (remote control, for synchronous operation) |
External input/output connectors | External Trigger : TRG IN, TRG OUT External Clock : CLK IN, CLK OUT (can be output at any divide-by-difference ratio) |
External Device Interface | Keyboard I/F: USB mouse I/F: USB external display I/F: VGA connector (15-pin for monitor) USB I/F: USB 2.0 (3 ports on the front) or USB
3.0 (2 ports on the back) * The USB 2.0 port on the front is limited to about 100mA of current that can be supplied. |
power | AC 100~240V, DC 10~30V (AC/DC dual) Built-in battery for instantaneous power failure |
Current consumption | Approx. 1.8A (when 100 AC 40V, CDV-8B sheets are mounted) |
Operating temperature and humidity range | 0~40°C,20~80%(non-condensing) |
Storage temperature range | -20~60℃ |
Vibration resistance | [EDX-5000A-64-H,EDX-5000A-64-S] 49.0m/s2(5G),5~55Hz(non-operating) 29.4m/s2(3G),5~55Hz(operating)[EDX-5000A-64-H] 9.8m/s2(1G),10~200Hz(operating) [EDX-5000A-64-S] |
Impact resistance | 196.1m/s2(20G),11ms |
External dimensions | 365(W)×159(H)×300(D)mm not including protrusions |
mass | Approx. 11.5kg (main unit only) |
Synchronous operation | Multiple EDX-5000A units can be connected to perform simultaneous acquisition operations. EDX-5000A and EDX-3000B, EDX-5000A and EDX-200A can be combinedNumber of synchronous connections: 10 (all EDX-5000A-80: up to 800 channels), but up to 200 when combined with EDX-8A |
Optional features | CAN data acquisition (up to 512 input channels) GPS data acquisition (trajectory data acquisition, time synchronization function) |
Online control | Control by DCS-100A using LAN I/F |
Standard Accessories | AC cable P-18 (w/CM-52), DC power cable P-70, ground wire P-72, Simple instruction manual, CD-R (instruction manual), EDX accessory bag, AC fuse, DC fuse, soldering plug for DIO,plug shell case for DIO |
---|---|
Sold separately | Synchronous cable N-128 Remote Control Unit RCU-42A Belt hook dummy panel for EDX EDX1P-DUMMY EDX2000-DUMMY |
EDX-5000A Main Unit Software ETA-55A Specifications
Measurement conditions
Channel conditions | Measurement ON/OFF, Measurement mode, Range, High-pass filter, Low-pass filter, Digital high-pass filter, Digital low-pass filter, Balance ON/OFF, Cal, calibration coefficient, offset, offset zero ON/OFF, unit, channel name, measurement range, rated capacity, rated output, upper check value, lower limit check value, number of numerical display digits (display items can be selected arbitrarily) Internal sensitivity registration ON/OFF, input cable, cable correction value |
---|---|
Number of data that can be acquired | Up to the remaining capacity of the internal disk |
Manual measurement | Acquire between REC and STOP or from REC to a specified number of data. |
Interval measurement | Acquisition is performed automatically by setting the acquisition start time and acquisition interval. |
Trigger measurement | Start/stop acquisition according to the set trigger conditions.
・Common trigger condition to AD format Analog trigger condition ・Digital trigger condition trigger bit: arbitrary 1-bit ・External trigger condition trigas rope: rise/fall ・Combined trigger condition from any 4 analog channels/digital channels, 1 external trigger channel, and 1 manual trigger channel AND/OR: Analog triggers, digital triggers, and external triggers can be logically determined by AND/OR Trigger level: analog channel set by physical quantity, digital channel 0,1 trigas rope: rise/fall |
TEDS Information Reading | Reading TEDS information and automatically setting according to the imported conditions |
Reading and saving measurement condition files | Import ME5 format files and save set measurement conditions based on the imported measurement conditions to ME5 format files *ME5 format files can also be read and saved from DCS-100A and EDX-3000B compatible software. |
Measurement behavior
Monitor measurement, acquisition start, acquisition stop, acquisition end, balance execution, CAL output, etc. are possible. | |
Real-time processing | The maximum processing sampling frequency limit that can be performed simultaneously with monitoring and acquiring measurement data is 10kHz. |
Video acquisition with a webcam
Cameras used | DirectX-enabled webcam (webcam recognized by the OS as an imaging device) |
---|---|
Number of cameras used | 1 unit |
Resolution Size | Up to 640×480 |
Frame Rate | Up to 30fps |
Save file format | AVI format * The resolution size and frame rate are determined by the webcam used. A webcam is required separately. |
Measurement conditions when acquiring video | |
Measurement mode | Manual, Manual (Number of Acquired Data Setting) |
Arithmetic processing
Digital Filters | Characteristics: Butterworth (IIR) Type: Low-pass filter, High-pass filter order: Amplitude ratio at 1~8th order cut-off point: -3dB Attenuation characteristics: -6 dB×N dB/oct. (N is the degree)* Can be used with low-pass filters with conditioner cards |
---|---|
Four arithmetic operations | Number of Calculation Channels: Maximum 64 Channel Calculation Channel Conditions: Arithmetic ON/OFF, Arithmetic Expression (up to 200 half-width characters), Unit, Number of digits displayed, Channel
name (up to 40 half-width characters), Zero Operation, Zero Operation, Zero-value operator/Constant Arithmetic function DTAN: Inverse Tangent (Return Value: Angle) EXP: Exponential HMX: Maximum principal strain HMN: Minimum principal strain |
FFT Analysis
Analysis Type | linear spectrum, power spectrum, cross spectrum, autocorrelation, cross-correlation |
---|---|
Window Functions | OFF,Humming,Hanning,Feiel,Blackman,Gaussian |
Number of analysis data | 256,512,1024,2048,4096,8192 |
Number of analysis channels | Up to 1 channels in one graph |
Storage format | KS2 file saved in Kyowa Standard File Format (KS2 format) Version: 01.06 |
---|
Monitor Screen
Time Series Graph | X-axis: Time, Y-axis: Up to 8 graphs of measured values (physical quantities) can be displayed. |
---|---|
X-Y Graph | X-axis: measured values, Y-axis: Graphs of measured values can be displayed in combinations of up to 4 channels |
Bar Graph | Up to 80 channels can be displayed. |
Yen Meter | Any one channel can be displayed with a circle meter |
Numeric display | Up to 80 channels can be displayedMaximum and minimum values for each channel can be displayed. |
FFT Graph | Up to 4 graph channels can be displayed to display FFT analysis results |
Webcam | View webcam capture images |
Input overdisplay | Numerical values, backgrounds, etc. of input overchannels can be displayed in red (excluding FFT graphs) |
Graph Scale | Autoscale can be displayed on the Y axis of a time series graph, the X and Y axes of an X-Y graph, and the Y axis of a Bar graph |
Cursor display | Time series graphs, X-Y graphs can display cursor positionsCursor positions can be displayed in physical quantities |
Number of simultaneous displays | Up to 12 graphs |
Image storage | Each graph can be saved as an image file |
Preferences
Synchronous operation settings | Choose from standalone, synchronous master, and synchronous slave |
---|
other
Oscillator switching (internal and external), operating beep, balance specification value, AD data format switching (16/24 bit), option switch settings |
GPS/multi-channel CAN optional functions
CAN Data Acquisition | Up to 1 channels of CAN data can be acquired per unit (CAN data is stored as E512A files. * However, CAN conditions are set with EDX-4B compatible software. |
---|---|
CAN Trigger | Analog triggers, digital triggers, and complex triggers are used to start/stop acquisition. * However, the conditions for the CAN trigger are set with EDX-3000B compatible software. |
Graph display | Numerical display, graph display at the same time as analog data |
Point Zero (manual measurement) | Time data received from GPS satellites can be used to start manual measurement at zero second time (time when ms is 0) |
Point Zero (Interval Measurement) | Time data received from GPS satellites can be used to initiate interval measurement at zero seconds (when ms is 0). |
GPS Data Acquisition | GPS data (latitude, longitude, speed, direction, etc.) can be monitored and acquiredGPS data is saved in NMEA file format |
EDX-3000B Compatible Software ELA-55A Specifications
Measurement conditions
Channel conditions | Measurement ON/OFF, Measurement mode, Range, High-pass filter, Low-pass filter, Digital high-pass filter, Digital low-pass filter, Balance ON/OFF, Cal, calibration coefficient, offset, offset zero ON/OFF, unit, channel name, measurement range, rated capacity, rated output, upper check value, lower limit check value, number of numerical display digits (display items can be selected arbitrarily) Internal sensitivity registration ON/OFF, input cable, cable correction value |
---|---|
Number of data that can be acquired | Up to the remaining capacity of the internal disk |
Manual measurement | Acquire between REC and STOP or from REC to a specified number of data. |
Interval measurement | Acquisition is performed automatically by setting the acquisition start time and acquisition interval. |
Trigger measurement | Start/stop acquisition according to the set trigger conditions.
・Common trigger condition to AD format Analog trigger condition ・Digital trigger condition trigger bit: arbitrary 1-bit ・External trigger condition trigas rope: rise/fall ・Composite trigger condition from any 4 analog channels/digital channels, 1 external trigger channel, and 1 manual trigger channel AND/OR: Analog triggers, digital triggers, and external triggers can be logically determined by AND/OR Trigger level: analog channel set by physical quantity, digital channel 0,1 trigas rope: rise/fall |
Saving and reading measurement condition files | Set measurement conditions can be saved and read |
Saving and reading sensor information files | Sensor information files (CSV format files) can be saved and read under channel conditions |
DIO Settings | 【Number of input/output points】 Up to 8 points [Input/output setting] Digital input/digital output/remote control input can be set for each bit |
TEDS Information Reading | Reading TEDS information and automatically setting according to the imported conditions |
Reading and saving measurement condition files | Read ME5 format files and save set measurement conditions based on the imported measurement conditions to ME5 format files * ME5 format files can also be read and saved from DCS-100A and the main unit software. |
Measurement behavior
Monitor measurement, acquisition start, acquisition stop, acquisition end, balance execution, CAL output, etc. are possible. | |
Real-time processing | The maximum processing sampling frequency limit that can be performed simultaneously with monitoring and acquiring measurement data is 10kHz. |
Video acquisition with a webcam
Cameras used | DirectX-enabled webcam (webcam recognized by the OS as an imaging device) |
---|---|
Number of cameras used | 1 unit |
Resolution Size | Up to 640×480 |
Frame Rate | Up to 30fps |
Save file format | AVI format * The resolution size and frame rate are determined by the webcam used. A webcam is required separately. |
Measurement conditions when acquiring video | |
Measurement mode | Manual, Manual (Number of Acquired Data Setting) |
Arithmetic processing
Digital Filters | Characteristics: Butterworth (IIR) Type: Low-pass filter, High-pass filter order: Amplitude ratio at 1~8th order cut-off point: -3dB Attenuation characteristics: -6 dB×N dB/oct. (N is the degree)* Can be used with low-pass filters with conditioner cards |
---|---|
Four arithmetic operations | Number of Calculation Channels: Maximum 64-channel operation channel conditions: Arithmetic ON/OFF, arithmetic expression (up to 200 half-width characters), unit, number of digits displayed, channel name (up to 40 half-width characters), Zero operation, Zero-value
operator/constant arithmetic function DTAN: Inverse Tangent (Return Value: Angle) EXP: Exponential HMX: Maximum principal strain HMN: Minimum principal strain |
FFT Analysis
Analysis Type | linear spectrum, power spectrum, cross spectrum, autocorrelation, cross-correlation |
---|---|
Window Functions | Square,Humming,Hanning,Feiel,Blackman,Gaussian |
Number of analysis data | 256,512,1024,2048,4096,8192 |
Number of analysis channels | Up to 1 channels and up to 4 windows can be displayed in one window |
preservation | Analysis results can be saved in FFT analysis result file format (CSV format) |
Storage format | KS2 file saved in Kyowa Standard File Format (KS2 format) Version: 01.06 |
---|
Monitor Screen
Time Series Graph | X axis: time, Y axis: Up to 16 channels of graphs of measured values (physical quantities) can be displayed1~1 graphs can be displayed in one window |
---|---|
Time series (all channels) graph | X-axis: Time, Y-axis: All measurement channels of graphs of measured values (physical quantities) can be displayed in the same color line. |
Time Series (DIV) Graph | X-axis: Time, Y-axis: Up to 16 graphs of measured values (physical quantities) can be displayedThe 0-point position of the display channel can be placed at any position on the Y-axis divider. |
X-Y Graph | X-axis: measured values, Y-axis: Graphs of measured values can be displayed in combinations of up to 8 channels |
Bar Graph | Up to 1 channels can be displayed in one graph, 32~1 graphs can be displayed in one windowPeak hold can be displayed (peak value can be displayed numerically) |
Digital Graph | X axis: time, Y axis: Bit data of any digital channel up to 16 bits can be displayed, 1~1 graphs can be displayed in one window |
Yen Meter | Any one channel can be displayed with a circle meter |
Bar Meter | Any single channel can be displayed with a vertical or horizontal meter |
Numeric display | Arbitrary 1-channel display, optional 16-channel display, All channel display (maximum and minimum values for each channel can be displayed) |
FFT Graph | Up to 4 graph channels can be displayed to display FFT analysis results |
Webcam | View webcam capture images |
Input overdisplay | Input over channel numbers can be displayed in red (except FFT graphs) |
Graph Scale | The Y axis of autoscale/full-scale displayable time series graphs can be switched by 1-axis, 2-axis, and channel on the Y axis of time series graphs, the X-Y axis of X-Y graphs, and the Y axis of Bar graphs. |
Screen display color | Can be changed arbitrarily on a window-by-window basis |
Titles,Labels | Title, X-axis, Y-axis label can be set arbitrarily |
Number of simultaneous displays | Numeric windows: Up to 8, Graph windows: Up to 8 (including FFT windows) Up to 16 numeric and graph windows can be displayed |
Witness lines | Arbitrary auxiliary lines can be displayed on the X and Y axes of time series graphs, X-Y graphs, and Bar graphs (up to 4 auxiliary lines can be displayed for each X and Y axes). |
Comparison data display | Comparison data (KS2 data acquired in the past) can be displayed in time series graphs (all channels, excluding DIV) and X-Y graphs can be displayed to compare
the monitor data with the previous data. |
Preferences
Synchronous operation settings | Choose from standalone, synchronous master, and synchronous slave |
---|---|
Control mode | Switching offline and online |
Automatic file conversion | File conversion (CSV format, XLS format, XLSX format, RPC III format) is automatically performed at the end of the measurement. |
other
Oscillator switching (internal and external), operating beep, balance specification value, AD data format switching (16/24 bit), option switch settings |
GPS/multi-channel CAN optional functions
CAN Data Acquisition | Up to 1 channels of CAN data can be acquired per unit (CAN data is stored as E512A files. ) |
---|---|
CAN Trigger | ・Analog trigger condition trigger channel: Configurable from any CAN channel Trigger level: Set by CAN data Trigas rope: Rise/fall/ digital trigger condition Trigger channel: Any 1-bit trigger level of any CAN channel: 0, 1 ・ Composite trigger condition AND/OR: Analog trigger, digital trigger can be logically determined |
Communication conditions | Communication standard, communication speed, termination resistor * Sample point, number of samples, and resynchronization jump width are automatically set. |
Measurement ID condition | Port number, format, frame ID, data length |
Measurement channel conditions | Start bit, bit length, data type, endianness, calibration factor, offset (conditions for extracting CAN data and converting it to physical quantities) |
Graph display | Numerical display, graph display at the same time as analog data |
Reading CANdb Files | File Format: DBC Format Max Network Nodes: 1024 Max Messages: 2048 Max Signals: 4096 |
Point Zero (manual measurement) | 1Using the time data received from GPS satellites, it is possible to start manual measurement at zero second time (time when ms is 0) |
Interval (GPS sync) measurement | Time data received from GPS satellites can be used to initiate interval measurement at zero seconds (when ms is 0). |
GPS Data Acquisition | GPS data (latitude, longitude, speed, direction, etc.) can be monitored and acquiredGPS data is saved in NMEA file format |
Data Playback and Analysis Software Specifications
Data playback
Graph display | 4 patterns of display conditions can be set, displayed in one of 1, 2, or 4 graphs |
---|---|
Y-Time Graph | Up to 1 channels can be displayed in one graph, autoscale on the Y axis |
X-Y Graph | 1Arbitrary 4 channels can be displayed on both the X axis and the Y axis fixed graph, and both the X axis and the Y axis can be displayed autoscale |
Full data display | All data can be displayed in units of 4 channels |
Numerical data display | List numerical data of any 16 channels |
Cursor display | Numerical display of cursor positionEnlarged display between two cursorsMaximum and minimum values between two cursors can be displayed |
scroll | Y-Time graph scrolls the X-axis |
Data file editing | Arbitrary data range and arbitrary channel extraction of data fileconvertible data file title , comment, channel conditions can be displayed and editable |
MAX/MIN data can be displayed | Maxima and minimum values for each channel can be displayed (up to 5 units) 400 data before and after centered on local maximums and micromas can be displayed graphically |
KS2 Files | Block number support (1 block display, all block display) voice data list and voice data playback |
Playback operations | Playback, reverse playback, continuous playback, frame forwarding, reverse frame forwarding, moving to playback start position and playback end position, setting playback start position and playback end position, playback speed (0.1~20x speed) measurement data, video, GPS data can be played at the same time |
AVI Image Files | Playback frame rate setting, playback start frame number (time) setting |
analysis
Viewing Acquired Data
Statistical operations | List maximum, minimum, mean, and standard deviation of arbitrary data ranges, and save statistical calculation results to CSV format file. |
---|---|
Four arithmetic operations | Calculation can be performed on channels in up to 2 data files and the calculation result can be saved in a new data fileUp to 200 expressions (up to 320 half-width characters) can be setInputpossibleoperators/constant additions: +, multiplication: *, division: /, exponentiation: ^, pi: PI, parentheses: () Arithmetic function (return value: radians) LOG: Common Logarithm HMN: Minimum principal strain |
FFT Analysis
Analysis Type | linear spectrum, power spectrum, cross spectrum, autocorrelation, cross-correlation, coherence, transfer function |
---|---|
Number of analysis data | 256,512,1024,2048,4096,8192,16384,32768 |
Window Functions | OFF,Humming,Hanning,Feiel,Blackman,Gaussian |
filter | 1 stages of 2,5,10,20,50,100,200,500,1000,2000,12Hz and FLAT |
Number of integrations | 0:None,1~2 |
Average number of times | 0~(0: Entire waveform) |
Number of shift data | 2 or more |
Analysis result graph display | 【Linear Spectrum】 Time Series Graph: Possible Analysis Graph 1: Amplitude (Linear)/Amplitude (Logarithmic) Analysis Graph 2: Phase[Power Spectrum] Time Series Graph: Possible Analysis Graph 1: Amplitude (Linear)/Amplitude (Logarithmic) Analysis Graph 2: – [Cross Spectrum] Time Series Graph : Possible [Autocorrelation] Time Series Graph: Possible Analysis Graph 1: Correlation Cross-correlation] Analysis Graph 1: Correlation 【Coherence】 【Transfer Function】 Analysis Results in CSV File |
Frequency analysis
Analysis Type | Maximum/minimum method, maximum/minimum method, 1D rainflow method, amplitude method, 1D time rate method, 1D rainflow method + local maximum/minimum method 1D rainflow method + maximum/minimum method 2D rainflow method |
---|---|
Number of slices | 1D type: 10 (±5) ~ 256 (±128) even numbers 2D type: 10~50 even
number slice width, hysteresis, offset (during maximum/minimum value method), etc. can be set. |
Analysis result graph display | Tabulation display, drawing display (2D display in the case of 3D rainflow method) |
Life Prediction Processing | Life prediction is performed from frequency analysis results obtained by the 1D rainflow method, 2D rainflow method, and amplitude method, and the results can be displayed and saved to a file (S-N data files can be read). |
Filtering | Digital Filter: IIR Filter Characteristics: 4th Order Butterworth (Cutoff: -6dB, No Phase Delay) High-Pass Lowpass Filter : FLAT ~ 500kHz (Up to 1/2 of the sampling frequency can be set)Mirroring analysis results are as follows: Can be saved in additional formats |
Differentiation and Integration | Number of differentiations and integrations 0: None, 1~2 Analysis results can be saved in additional formats |
other
Batch conversion of multiple data files | CSV format file, Excel🄬Format file (XLS,XLSX),RPCIII. format file |
---|---|
Data File Merge | Combines multiple data files acquired by synchronous operation (master/slave) into a single file |
Reverse file conversion | Convert files converted to CSV format files to KS2 files |
File splitting | Extract a multi-block data file with only the specified block number data into one file. |
Batch analysis processing | Analysis processing (frequency analysis, filtering, calculus processing) of multiple data files under the same conditions is possible. |
Overwrite display of multiple files | Up to 16 data files can be overwritten as time series data, and the display start position can be set arbitrarily for each data file |
CSV format conversion of multiple KS2 files | Multiple KS2 files acquired by interval measurement, etc. can be converted into a single CSV format file |
Dimensions
Remarks
If you want to install EGPC-50A additionally, you need to pick it up from us.