Japan, Kyowa Electronic Instruments Co., Ltd.
Kyowa UCAM-60C All-in-One Data Logger UCAM-60C/65C Data Logger
Made in Japan
Manufacturer: Kyowa Electronic Instruments Co., Ltd.
Model: UCAM-60C All-in-One Data Logger
Long-term measurement data stored in the internal memory (built-in memory approx. 1.8GB)
Stored data can be easily recovered
by USB memory
Up to 1000 channels of measurement (when connected to an external scanner USB-70B series)
0.1×10-6High resolution of strain, 20000×10-6Strain measurement (4 gauge method)
All-in-one data logger (UCAM-60C)
Online dedicated data logger (UCAM-65C)
Feature
The UCAM-60C is an all-in-one measuring instrument designed for ease of use. Equipped with easy-to-operate keys, easy-to-understand Japanese display, bright and easy-to-read display, and a printer that allows you to quickly see measurement results, it is equipped with functions necessary for on-site measurement.
Specification
Model Name | UCAM-60C-AC: AC Power Supply, UCS-60B Sold Separately, Operation Key, Display, Printer Available UCAM-60C-DC: DC Power Supply, UCS-60B Sold Separately, Operation Key, Display Unit, Printer Available |
---|---|
Measurement object | Strain gauge, strain gauge transducer (TEDS compatible product can be connected), DC voltage, DC current, civil engineering transducer with temperature detector, potentiometer type sensor, temperature (thermocouple, platinum resistance thermometer detector) |
Connectable Scanner | USS-61B, 62B, 63B (dedicated scanning unit, mounted on the top of the main unit) Using the optional scanner interface (USI-67A), the following scanners can be connected. USB-70B series (using USI-67A) |
Number of input channels | When dedicated scanning unit is installed: up to 30 When an external scanner is connected: up to 1000 Dedicated scanning unit + external scanner up to 1000 |
Input terminal | Soldered/screwed shared-type terminal block (screws: M3 with washer) NDIS4102 (7-pin) connector (when using USS-62B scanning unit) |
Channel switching element | Semiconductor relay |
Scanning Speed | 50ms/channel (standard mode) 0.28s/channel (high resolution mode) 20ms/channel (high-speed mode)
* High-resolution mode and high-speed mode can only be set with a dedicated scanning unit High-resolution : Strain 4 gauge method and voltage, current, and potentiometer sensors only |
Mode of operation | Real-time measurement monitor measurement, automatic measurement (interval measurement ) |
Measurement Functions | Initial value measurement (save measured value as initial value) Measure value measurement (measured value = original value – initial value) Original value measurement (without subtraction of initial value) Easy measure value measurement (auto-zero function enabled) (measured value = original value – zero value)
* Batch configuration for all channels |
Coefficient calculation function | Multiplication function such as calibration coefficients, physical quantity conversion of calibration measured values by TEDS, scaling and correction are possible. |
unit | 59 types |
Automatic measurement function
Interval measurement | Automatic measurement at set time intervals |
---|---|
Trigger measurement | Trigger start by relative value (constant change) or absolute value |
Trigger interval measurement | Combination of the above trigger measurement and interval measurement |
Storage features | Built-in memory capacity: about 1.8GB |
---|
Strain measurement (standard mode)
Constant voltage bridge power supply | DC approx. 2V or DC approx. 5V |
---|---|
Constant Current Bridge Power Supply | Approx. 5.7mA (at 350Ω bridge resistance) Approx. 16.7mA (at 120Ω bridge resistance) |
Scanning Speed | 50ms/channel |
Gauge rate | Fixed 2.00 (can be corrected by 2.00/Ks by coefficient calculation function) |
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | Measuring range: 0~±50000×10−6Strain resolution: 1×10−6Strain accuracy: ± (0.05%+1 of display value)×10−6Strainmeasurement range: ±50000~±500000×10 −6Strain |
Strain measurement (high resolution mode)
Constant voltage bridge power supply | DC approx. 5V |
---|---|
Constant Current Bridge Power Supply | Approx. 16.7 mA (at 350 Ω bridge resistance) |
Scanning Speed | 0.28s/channel |
Gauge rate | Fixed 2.00 (can be corrected by 2.00/Ks by coefficient calculation function) |
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | Measuring range: 0~±20000×10−6Strain resolution: 0.1×10−6Strain accuracy: ± (0.05% of display value + 0.3)×10−6Strainmeasurement range: ±20000~±200000×10 −6Strain Measuring range 0~±15000×10 -6Strain |
Strain measurement (high-speed mode)
Constant voltage bridge power supply | DC approx. 2V |
---|---|
Constant Current Bridge Power Supply | Approx. 5.7mA (at 350Ω bridge resistance) Approx. 16.7mA (at 120Ω bridge resistance) |
Scanning Speed | 20ms/channel |
Gauge rate | Fixed 2.00 (can be corrected by 2.00/Ks by coefficient calculation function) |
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | Measuring range: 0~±50000×10−6Strain resolution: 1×10−6Strain accuracy: (0.08%+3 of display value)×10−6Strainmeasurement range: ±50000~±500000×10 −6Strain |
Voltage measurement (standard mode)
Scanning Speed | 50ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | CH mode: V/500mV Measurement range: 0~±50.000mV Resolution: 1μV Accuracy: ± (0.05% of display value + 0.003) mV Input resistance: 10MΩ or more
CH mode: V/500mV Measurement range: ±50.00~±500.00mV Resolution: 10μV Accuracy: ± (0.05% of display value + 0.03)mV Input resistance: 10MΩ or more CH mode: V/50V Measuring range: 0~±5.0000V Resolution: 100μV Accuracy: ± (0.05% of display value + 0.0002)V Input resistance: 1MΩ or more CH mode: V Accuracy: ± (0.05% of the displayed value + 0.002) V * The auto-ranging function automatically switches the resolution and accuracy according to the measurement range. |
Voltage measurement (fast mode)
Scanning Speed | 20 ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | CH mode: V/500mV Measurement range: 0~±50.000mV Resolution: 1μV Accuracy: ± (0.08% of display value + 0.006)mV Input resistance: 10MΩ or more CH mode: V/500mVMeasurement range: ±50.00~±500.00mV Resolution: 10μV Accuracy: ± (0.08% of display value + 0.06) mV Input resistance: 10MΩ or more CH mode: V/50V Measuring range: 0~±5.0000V Resolution: 100μV Accuracy: ± (0.08% of display value + 0.0006)V Input resistance: 1MΩ or more CH mode: V Accuracy: ± (0.08% of the displayed value + 0.006) V * The auto-ranging function automatically switches the resolution and accuracy according to the measurement range |
Current measurement (standard mode)
Scanning Speed | 50ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | CH Mode: I/50mA Measurement Range: 0~±50.00mA Resolution: 10μA Accuracy: ± (0.05% of the displayed value + 0.01)mA
(Note) External shunt resistor 250 Ω (high accuracy) required |
Current measurement (fast mode)
Scanning Speed | 20ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Measuring range, resolution, accuracy | CH mode: I/50mA Measurement range: 0~±50.00mA Resolution: 10μA Accuracy: ± (0.08% of the displayed value + 0.01) mA
* Only the dedicated scanning unit can be set (Note) External shunt resistance 250 Ω (high accuracy) is required |
Thermocouple Temperature Measurement (Standard Mode)
Scanning Speed | 50ms/channel |
---|---|
measuring range, resolution, accuracy | [Type: K] Measuring range: -200.0 ~ 1230.0°C Resolution: 0.1°C Accuracy: ±0.7°C Internal reference contact compensator accuracy: ±0.5°C (when input terminal temperature is balanced) (ambient temperature range 0~50°C) [Type: T] Measuring range: -200.0 ~ 400.0°C Resolution: 0.1°C Accuracy: ±0.7°C Internal reference contact compensator accuracy: ±0.5°C (when input terminal temperature is balanced) (ambient temperature range 0~50°C ) [Type: E] ) 【Type J】 ) [Type: R] ) * Accuracy does not |
Civil engineering transmitter with temperature measurement function temperature measurement (standard mode)
Scanning Speed | 50ms/channel |
---|---|
Measuring range, resolution, accuracy | Measurement range: -50.0~200.0°C Resolution: 0.1°C Accuracy: ±0.5°C
* Target physical quantity and temperature can be measured on one channel. |
Platinum resistance thermometer detector temperature measurement (standard mode)
Scanning Speed | 50ms/channel |
---|---|
measuring range, resolution, accuracy | [Type: Pt100] Measuring range: -200.0~660.0°C Resolution: 0.1°C Accuracy: ±0.3°C [Type: JPt100] Measuring range: -200.0~510.0°C Resolution: 0.1°CAccuracy: ±0.3°C *Termination system: 3-conductor type |
Potentiometer sensor measurement (standard mode)
Scanning Speed | 50ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Sensor power supply | DC approx. 2V |
Potentiometer resistance | 1k~10k Ω |
measuring range, resolution, accuracy | Channel mode: POT. Measuring range: 0~±50.00% Resolution: 0.01% Accuracy: ±0.1%FS |
Potentiometer sensor measurement (high-speed mode)
Scanning Speed | 20ms/channel |
---|---|
Initial value storage range | Same as measuring range |
Sensor power supply | DC approx. 2V |
Potentiometer resistance | 1k~10k Ω |
measuring range, resolution, accuracy | Channel mode: POT. Measuring range: 0~±50.00% Resolution: 0.01% Accuracy: ±0.1%FS* Only dedicated scanning unit can be set |
Built-in clock | Built-in real-time clock (battery backup) |
---|---|
indication | VFD display128 ×64-bit |
printer | Printing method : Thermal paper width: 58 mm (24 digits) Printing speed: 60 mm/sec (max.) |
interface | RS-232C LAN (10BASE-T/100BASE-TX) USB 2.0 (for collecting acquisition data from the built-in memory using a USB memory)*1 Data cannot be saved directly to a USB memory *2 USB memory Capacity: Up to 32GB File format: FAT32 *3 Recommended USB memory: GH-UFI-XSD2G (Manufacturer: Green House) |
File conversion function | Convert files saved in binary format to CSV format |
Self-diagnostic function | Checking displays, printers, bridge power supplies, input disconnection, input/output resistance, insulation resistance, mode detection, etc. |
TEDS compliant | Interface IEEE1451.4 Mixed mode Transducer Interface Class 2 compatiblesensor Sensor with information according to IEEE template No. 33, cable length 30m or less * When the dedicated scanning unit USS-61B/62B/63B is installed |
Operating temperature and humidity range | Temperature 0~50°CHumidity 20~85%(non-condensing) |
Storage temperature range | -20~60℃ |
Maintenance settings function | ACOM of measurement circuit can be switched to floating or GND connection |
power | AC 100~240V (AC power supply dedicated unit) DC 10~16V (DC power supply dedicated unit) * DC power supply dedicated unit has power control function |
Current consumption | AC 100 V: 0.5 A or less (with 3 dedicated scanning units) 12 VDC : 4 A or less (with 3 dedicated scanning units) |
External dimensions | 360(W)×88(H)×400(D)mm (excluding protrusions) |
mass | Approx. 6.3 kg (without dedicated scanning unit) Approx. 9.6 kg (with 62 dedicated scanning units USS-3B) |
Standard Accessories | AC power cable P-18 (with 2-pin conversion plug CM-52) (AC power supply dedicated unit) DC power cable P-76 (DC power supply dedicated unit) Record paper UCAM-60A-RP (1 volume) Driver fuse CD-R (instruction manual)Ground wire P-72 |
---|---|
Sold separately | Record Paper UCAM-60A-RP (1 box of 10 rolls) |
Dimensions
Configuration
[Data logger (main unit)]
UCAM-60C-AC (AC power supply, control software sold separately, operation key, display, printer installed)UCAM-60C-DC (DC power supply, control software sold separately, operation key, display unit, printer installed)
UCAM-65C-AC (AC power supply, control software included, PC control type)
UCAM-65C-AC-0 (AC Power Supply, Control Software Sold Separately, PC Control Type) UCAM-65C-DC (DC Power Supply, Control Software Included, PC Control Type) UCAM-65C-DC-0 (DC Power Supply, Control Software Sold Separately, PC Control Type)
[ Dedicated scanning unit (sold separately) ]
USS-61B (TEDS compatible) USS-62B (with NDIS4102 (7-pin) connector, TEDS compatible) (*1) USS-63B (for civil engineering measurement, with lightning surge protector, TEDS compatible)
* One dedicated scanning unit can measure 1 channels Up to 10 dedicated scanning units can be mounted on the main unit [External scanner] Using the separately sold scanner interface, the following scanners manufactured by our company can be connected.
USB-3B ( USI-70A [Scanner Interface]
For USI-67A USB-67B [External input/output unit]
UIO-70A 【 Control Software]
UCS-60B
Related Products
-