Japan, Kyowa Electronic Instruments Co., Ltd.
Kyowa USM-51B Strain/Voltage/Thermocouple Unit UCAM-550A High Speed Data Logger
Made in Japan
Manufacturer: Kyowa Electronic Instruments Co., Ltd.
Model: USM-51B Strain/Voltage/Thermocouple Unit
- Simultaneous sampling of all channels
- Continuous and simultaneous measurement of up to 50 channels at 1000 times / second
- 5 types of measurement units are available
Feature
A high-speed data logger capable of repeatedly measuring up to 1000 channels at 0.02 second intervals. Since it can be measured simultaneously at high speed, it can handle a wide range of phenomena from static phenomena to quasi-dynamic phenomena.
Specification
Input terminal | NDIS4102 (7-pin) connector + screw solder terminal block |
---|---|
Number of input channels | 10 |
Measurement object | 1 gauge method (built-in 2-wire/3-wire switching circuit): 120/350Ω 2 gauge method, 4 gauge method: 120 ~ 1000Ω Strain gauge transducer: 120 ~ 1000Ω potentiometer type sensor: 1k ~ 10kΩVoltage: ±20V thermocouple: K, T, E, J, R, N *Thermocouple resistance 1 kΩ or less |
Bridge Power Supply | Constant voltage DC 2V |
Potentiometer drive power supply | Constant voltage DC 2V |
Gauge rate | 2.00 Fixed |
Response frequency | DC~7.8Hz deviation: 0.5dB~-3.5dB (excluding thermocouples) |
Burnout check | Perform burnout only when checked |
TEDS | Reading information from sensors with built-in TEDS |
Strain/potentiometer sensor/voltage | (Strain range L) Measuring range: 0~±19000×10-6Strain resolution: 1×10-6Strain accuracy: ±0.08%FS(strain range H) Measuring range: 0~±300000×10 -6Strain (voltage) Measuring range: 0~±20V Resolution: 1mV Accuracy: ±0.08%FS |
thermocouple | (Thermocouple K) Measurement range: -200.0~+1200.0°C Resolution: 0.1°C Measurement accuracy: [When external reference contact is compensated] <When the input terminal temperature is balanced<b14>, the ambient temperature range: 25±10°C> -200.0~-100.0°C: ± (0.3%) rdg+0.8°C) -100.0~+1200.0°C:±(0.2%rdg+0.6°C) <When the input terminal temperature is balanced, the ambient temperature range: 25±10°C is not > -200.0~-100.0°C: ±(0.3%rdg+0.8°C) -100.0~+1200.0°C: ±( 0.2%rdg+0.6°C) [When internal reference contact compensation] < Input terminal temperature equilibrium, ambient temperature range: > Measuring range: -200.0~+350.0°C Resolution: 0.1°C Measurement accuracy: [When external reference contact is compensated] : -100.0~+350.0°C: ± (0.2%rdg+0.6°C) [When internal reference contact is compensated] <Input terminal temperature balanced<b130>, Ambient temperature range: 25±10°C> -100.0~+350.0°C: ±(0.2%rdg+1.6°C) <when the input terminal temperature is balanced, the ambient temperature range: 25± when the temperature is other than 10°C> -200.0~-100.0°C: ±(0.3%rdg+2.8°C) -100.0~+350.0°C:±(0.2%rdg+2.6°C) ( > -100.0~+800.0°C:±(0.2%rdg+0.6°C)【When internal reference contact compensation】 <Input terminal temperature balanced, ambient temperature range: 25±10°C> (Thermocouple J) > <When the input terminal temperature is balanced, the ambient temperature range: ±> -100.0~+750.0°C:±(0.2%rdg+1.6°C) <When the input terminal temperature is balanced, the ambient temperature range: ±> -100.0~+750.0°C:±(0.2% rdg +2.6°C) (thermocouple R) Measuring range: 0.0~+1600.0°C Resolution: 0.1°C Measurement accuracy: <When the input terminal temperature is balanced, the ambient temperature range: 25±When the temperature is other than 10°C> 0.0~+Below 100.0°C: ±(0.6%rdg+1.2°C) +100.0~+1600.0°C:±(0.5%rdg+1.0°C) <Input terminal temperature balanced, Ambient temperature range: 25±10°C> Less than 0.0~+100.0°C: ±( +100.0~+1600.0°C:±(0.5%rdg+3.0°C) (Thermocouple N) Measurement accuracy: [When external reference contact is compensated] < input endAt child temperature equilibrium, ambient temperature range: 25±10°C> -200.0~-100.0°C: ±(0.3%rdg+0.8°C) -100.0~+1250.0°C:±(0.2%rdg+0.6°C) <when input terminal temperature is balanced, ambient temperature range: 25±10°C> -200.0~-100.0°C: ±(0.3%rdg+0.8°C) -100.0~+1250.0°C:±(0.2%rdg+0.6°C) <Input terminal temperature equilibrium, Ambient temperature range: ±> |
Sold separately | terminal cover UM-51B |
---|