Japan, Tokyo Keiso
Tokyo Keiso MA-920 Intelligent, Pure Electronics Micro Flowmeter
Made in Japan
Manufacturer: Tokyo Keiso
Model: MA-920
MA-920 MICRO FLOWMETER is a metal tube variable area flow meter which has local indication with transmitting of minute flow rate of liquids and gases.
TOKYO KEISO’s long time production know-how and recent electronics technology have been successfully combined.
The existing micro flow meters generally need a signal linearizer due to mechanical problem of very minute sensing part.
In MA-920, integrated microprocessor takes care of these automatic compensation based on individual stored calibration data and achieves high accuracy even for small flow rate.
The 4 to 20 mA with 2-wire system makes field wiring easier.
Features
- Covers very low flow rates of 0.6 to 3 L/h
- 2-wire 4 to 20 mA DC output
- Magnetic field sensor detects the float movement with electronics to eliminate hysteresis. High accuracy and repeatability have been achieved
- Easy- to- read digital LED display
- No liquid dampers are needed even for gas measurement application
- Ex d IIC T6 flame proof construction suitable even for Hydrogen atmosphere
Specifications
Measuring fluid | Liquids and Gases Viscosity limit for liquid flow measurement (Free from solids and particles) Meter size 1/2 Viscosity (Max.) 2.0mPa·s Meter size 3/4, 1 Viscosity (Max.) 5.0mPa·s |
---|---|
Measuring range | Liquid (Water) *Measuring range is subject to liquid viscosity. Max. 60 to 600 L/h Min. 0.6 to 3 L/h Gas (Air, 0°C, 1atm) Max. 2.2 to 22 m³/h (nor) Min. 10 to 100 L/h (nor) |
Fluid temperature | 0 to 120°C |
Fluid pressure | Max. 2.94MPa High pressure type19.6MPa *Material of high pressure type is SUS 316. Allowable pressure is subject to the flange rating when the connection is flange type. |
Material | Body: SCS14 Tapered tube: SUS316 Float: SUS316 PPS resin / Titanium will be used for 1/2″ meter size, and PPS resin / SUS316 will be used for 3/4 and 1″ meter sizes in gas measurement applications. |
Process connection | Screw (1/4, 3/8, 1/2, 3/4 or 1″) JIS10KFF flange (10A, 15A, 20A, or 25A) Opt. NPT or other screw *Other flanges than JIS10KFF |
Accuracy (Indication and output) |
10 L/h or more of F.S. (water) ±2%F.S. Less than 10 L/h of F.S. (water) ±3%F.S. 100 L/h (nor) or more of F.S. (air) ±2%F.S. |
Range ability | 10:1 (Accuracy guaranteed range) *Range ability of the meter with the full scale range smaller than 5L/min (water)is 10:2. *It may differ depending on the liquid viscosity. |
Output signal | 4 to 20 mA DC (2-wire system) Allowable load resistance: 600Ω or less (at 24 V DC, including wiring resistance) |
Power source | 24 V DC ±10% (Operating voltage range: 12 to 33 V DC) |
Indication | Upper (main display): Four digits for measurements and error codes Lower (sub display): Five digits for supplementary information |
Enclosure |
|