Japan, Kyowa Electronic Instruments Co., Ltd.
Kyowa PAA-5KR PAA-R Current-output Pressure Transducer
Made in Japan
Manufacturer: Kyowa Electronic Instruments Co., Ltd.
Model: PAA-5KR
The PAA-R is a strain-gage pressure transducer with a built-in amplifier and output signals from 4 to 20 mA. The cable is integrated with the PAA-R. It is suitable for inductrial equipment or pressure control system.
Features
• Strong against noise
• High reliability
• High safe overload 200%
Specifications
Performance
Rated Capacity | 500 kPa |
---|---|
Nonlinearity | Within ±0.2% RO |
Hysteresis | Within ±0.2% RO |
Current Output | 4 to 20 mA |
Environmental Characteristics
Safe Temperature | -20 to 80ºC |
---|---|
Compensated Temperature | -20 to 70ºC |
Temperature Effect on Zero | Within ±0.03% RO/ºC |
Temperature Effect on Output | Within ±0.02%/ºC |
Electrical Characteristics
SN Ratio | 50 dB or more |
---|---|
Load Resistance | 0 to 500 Ω |
Cutoff Frequencies of AMP | 1 kHz (Amplitude ratio at cutoff point -3 ±1 dB) |
Power Supply | 24 VDC (21 to 30 V), 30 mA or less |
Cable | 4-conductor (0.18 mm2) vinyl shielded cable, 4.6 mm diameter by 3 m long, bared at the tip (Shield wire is not connected to the case.) |
Mechanical Properties
Safe Overloads | 200% |
---|---|
Materials | Case: SUS (Metallic finish) Liquid-contacting part: SUS 630 |
Weight | Approx. 200 g (Excluding cable) |
Degree of Protection | IP64 (IEC 60529) |
Mounting Screw | G3/8, male |
Standard Accessories | Gasket (Mild copper) |
---|
Dimensions
To Ensure Safe Usage
To Ensure Safe Usage
When using the UCAM-60C M14 in voltage mode,
use a 250 Ω resistor to convert data to voltage.