- Home
- Products
- Kyowa EDB-41B EDX-100A/200A Battery Unit for Momentary Shutdown EDX-200A Universal Recorder
Japan, Kyowa Electronic Instruments Co., Ltd.
Kyowa EDB-41B EDX-100A/200A Battery Unit for Momentary Shutdown EDX-200A Universal Recorder
Made in Japan
Manufacturer: Kyowa Electronic Instruments Co., Ltd.
Model: EDB-41B EDX-100A/200A Battery Unit for Momentary Shutdown
- Equipped with a real-time digital filter 8th order digital filter can acquire beautiful waveforms.
High-speed/low-speed dual sampling Fast and slow phenomena can be measured with a reduced amount of data.
A high-speed sampling of all channels at 10kHz simultaneously (at 32 channels)Simultaneous measurement of 100 channels at up to 3kHz is possible
Can be used in high and low-temperature environments (EDX-200A-4T only) Supports -20°C~65°C - One-wire synchronization (other than EDX-200A-1)
Up to 8 units, can be used for large-scale measurement with distributed arrangement.
Feature
A compact and high-end EDX that supports a wide variety of measurements with a wide selection of conditioner cards while maintaining the convenience of the EDX-100A.
Specification
Compatible Models | EDX-100A-1,EDX-100A-1H EDX-100A-2,EDX-100A-2H EDX-100A-4,EDX-100A-4H EDX-200A-2H EDX-200A-4H |
---|---|
Built-in battery | Battery Type : Ni-MH Nominal Capacity : 730mAh, Nominal Voltage : 12V |
External power input | DC 11V~18V “DC IN” connector type: RM12BRD-4PH (Hirose) DC power supply or EDX-100A AC adapter (sold separately) |
Power output | When driving with an external power supply… External power supply voltage When driving with this unit (momentary stopping) … Approx. 11~15V “DC OUT” connector model : RM12BRD-4S (Hirose) |
How to Charge | Automatic charging starts with external power connection (charging time: up to 3.5 hours) Start discharging with refresh button ON (refresh time: up to 6.5 hours) |
Indicator | BATTERY LEVEL LED (SIMPLE REMAINING AMOUNT INDICATION) CHARGE (CHARGE/DISCHARGE INDICATION) |
buzzer | Alarm by buzzer sound during momentary pause |
Operating temperature range | 0~50°C(0~30°C when charging or refreshing) |
Operating humidity range | 20~90%(non-condensing) |
Storage temperature range | -20~50℃ |
External dimensions | 25(W)×132.5(H)×255(D)mm (excluding protrusions) |
mass | Approx. 500g |
Backup time (reference value) | Approx. 30 minutes EDX-100A-1(H) with 40 CDV-1B (8 channels), approx. 15 minutes at full load EDX-100A-2(H) with two CDV-40Bs (2 channels), approx. 16 minutes at full load EDX-5A-100(H) with 4 CDV-40Bs (4 channels) at
full load full load* when the built-in battery is fully charged, and the ambient temperature is 32~20°C |
Remarks
EDX-200A-2H/4H: CE compatible products can also be manufactured. Please contact us.