Japan, Magnescale
Magnescale MG80-SC / MG80-CM Interface
Made in Japan
Manufacturer: Magnescale Co., Ltd.
Model: MG80-SC / MG80-CM
Features
MG80-SC
- Main module with RS-232C port and Ethernet port as standard.
- Equipped with PLC link function compatible with PLCs from various companies.
(You can use RS-232C or Ethernet.. Please refer to the connection manual for compatible PLCs). - Up to 16 axes of length measuring units can be connected.
- Various settings of the unit can make from a PC by using the setting application for Windows PC.
MG80-CM
- Counter module for DK800S, DK and DT series (via MT13) digital gauges.
- DIN rail mount for easy installation
Specifications
Main module | |
---|---|
Model | MG80-SC |
Commnication protocol | RS-232C (2,400 bit/s to 230,400 bit/s) Ethernet (100BASE-TX:100Mbit/s) |
Function | Resolution selection (0.1 to 10 µm), reference point setting, reset, preset, master preset, current value / maximum value / minimum value / P-P value output of each axis, comparator function (Up to 8 sets of 2 or 4 steps can be set), PLC link function *1 |
Maximum connectable counter modules | Up to 16 MG80-CM can be connected to each MG80-SC. |
Mounting method | 35 mm DIN rail mounting |
Power supply | DC 10.8 to 26.4 V |
Power consumption | 2.4 W or less |
Maximum cable length * | RS-232C : 15 m *2 Ethernet : 20 m (shielded CAT5e recommend *3) |
*1 Refer to the connection manual for supported PLCs. *2 For RS-232C communication cable, please use DZ252 cable (sold separately). *3 The customer must provide the Ethernet communication cables. |
Counter module |
||
---|---|---|
Model | MG80-CM | |
Power consumption | 2.0 W or less (Measuring unit excluded) | |
Compatible measuring units | DK800S series, DK series, DT series (via MT13) | |
Alarm | Frequency response exceeded, Measuring unit not connected, Cable breakage |