Japan, Measuring Instrument
OSC92ET300 X-ray Fluorescence Coating Thickness Tester
Model: OSC92ET300
Features
– IBM /PC compatible PC system allows direct use of various applications of MS-Windows.
– In addition to the numerical filter, a mechanical filter has been introduced which allows testing under the best conditions with extremely accurate results.
– The minimum size collimator is 0.1φmm which allows measurements on the microscopic part of the sample.
– Mouse pointing on measuring position makes for a quick shoot of the target point. Consecutive automatic measurements can also be performed by simply entering the measuring position.
– The irradiated part is displayed on the Windows screen. The screen size changes automatically according to the size of using collimator
– With the self-check function, trouble-shooting is simple, clear and quick.
Specifications for each table type
Measurement table | Manual table | Automatic table | |||||||||||||
Head part | Size (mm) | 170 x 110 | 240 x 220 | ||||||||||||
Movement length | X (mm) | 70 | 200 | ||||||||||||
Y (mm) | 70 | 200 | |||||||||||||
Z (mm) | 80 | 50 | |||||||||||||
Test item’s height (Max.) | 80 (option 150) |
50 | |||||||||||||
Apparatus size (mm) | 602(W) x 463(D) x 732(H) | ||||||||||||||
Weight (kg) | 55 | 61 | |||||||||||||
Sample loading (kg) | 3 | 1 | |||||||||||||
Operation | Apparatus size (mm) | Main unit 182(W)x368(D)x372(H) / Monitor 412(W)x424(D)x420(H) | |||||||||||||
Weight (kg) | Main unit 9 / Monitor 21 | ||||||||||||||
Printer | Weight (kg) | 5.2 |
Specifications for all table types
OSC 92ET300 | |||||||||||||||
X-ray source | Immersed in oil, compact size Infinitesimal x-ray tube Target: Tungsten, Tube voltage 50KV, Tube current variable 1.2mA |
||||||||||||||
Irradiation | Vertical irradiation from above | ||||||||||||||
Detector | Proportional counter tube | ||||||||||||||
Collimeter | Five sizes (automatic change system) 0.1, 0.2, 0.5, 1.0, 2.0φmm (option : 0.05 x 0.5, 0.05φ) |
||||||||||||||
Measurable object | Atomic number 22 (Ti) ~82(Pb) | ||||||||||||||
Filter | Automatically changes numeric and mechanical filters (Ni and Co) filter | ||||||||||||||
Measurement range | Atomic No. 22-24: 0.2-approx. 20μm, Atomic No. 25-40: 0.1-approx. 30μm | ||||||||||||||
Atomic No. 41-51: 0.2-approx. 70μm, Atomic No. 52-82: 0.1-approx. 10μm | |||||||||||||||
Calibration curve | Auto-created, multi-point calibration | ||||||||||||||
Application | Single-layer, Two-layer, and Three-layer plating thickness test | ||||||||||||||
Simultaneous measurement of the thickness of composite component ratio | |||||||||||||||
Electroless Nickel plating thickness test | |||||||||||||||
Measurement function | Automatic measurement, Output mode setting, Automatic measurement condition | ||||||||||||||
setting, Spectrum measurement, Measuring the distance between two points |